Z-0011-2011

MAQUET HLM Tubing set with Bioline Coating (BEQ-TOP 21400 with 1/2 x 1/2 HG 0286 Connector; Device Part...

Maquet Cardiovascular Us Sales, LLC / initiated 2010-09-07 / Terminated / root cause: design

Maquet Inc. began a recall of MAQUET HLM Tubing set with Bioline Coating (BEQ-TOP 21400 with 1/2 x 1/2 HG 0286 Connector; Device Part Number BEQ-TOP 21400 on . The reason FDA records is "Tubing packs may contain defective 1/2 inch x 1/2 inch tubing connectors. Leaks at the shank of the connector may result in air aspiration in the venous line". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0011-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
MAQUET HLM Tubing set with Bioline Coating (BEQ-TOP 21400 with 1/2 x 1/2 HG 0286 Connector; Device Part Number BEQ-TOP 21400
Product code
DWE Tubing, Pump, Cardiopulmonary Bypass
Reason for recall
Tubing packs may contain defective 1/2 inch x 1/2 inch tubing connectors. Leaks at the shank of the connector may result in air aspiration in the venous line.
Root cause tag
design
FDA root cause
Device Design
Action
Maquet Getigne Group issued Product Recall/Urgent Medical Device Corrective Action letters dated September 7, 2010 to customers, identifying the affected product, the issue prompting the recall, and actions to be taken by customers.Customers are to check their inventory for affected product. Customers with affected product are to contact Laura Olilver at 201-995-8849 for an RMA number and shipment instructions. All firms are to coomplete and return the Corrective Action Response Form. Maquet Customer Service can bre contacted at 800-777-4222.
Quantity in commerce
22 items
Distribution
Worldwide Distribution: USA State of IL and country of the Netherlands.
Product codes and lots
Lot number 70052718 and 70048867