Z-0126-2011
EndoTool Drug Dose Calculator-Software Version 7.2.1800.3 and 7.2.1825.5
Abbott Medical / initiated 2010-07-26 / Terminated / root cause: software
Hospira, Inc. began a recall of EndoTool Drug Dose Calculator-Software Version 7.2.1800.3 and 7.2.1825.5 on . The reason FDA records is "The EndoTool software did not always provide an audible alert when a blood glucose measurement was due. This could result in a patient not receiving an appropriate dose of insulin". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0126-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- EndoTool Drug Dose Calculator-Software Version 7.2.1800.3 and 7.2.1825.5
- Product code
- DNN Calibrators, Ethyl Alcohol
- Reason for recall
- The EndoTool software did not always provide an audible alert when a blood glucose measurement was due. This could result in a patient not receiving an appropriate dose of insulin.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Hospira issued an Urgent Device Field Correction letter dated July 23, 2010 to consignees, providing details of the potential issue and the associated patient risk. Hospira is finalizing software enhancements which will resolve this situation within the next 90 days ,and will contact each consignee when it is available to be downloaded from the Hospira website. The notification letter also included a clinical bulletin, outlining mitigation instructions for clinicians to use until the software enhancement is available. Consignees were requested to complete and return an attached Reply Form by fax. Customers can contact the Hospira EndoTool support line at 1-877-349-4582 for clinical inquiries about this recall.
- Quantity in commerce
- 97 units total (96 version 7.2.1800.3 and 1 version 7.2.1825.5)
- Distribution
- Nationwide Distribution in states of CA, FL, IL, IN, KY, KS, MI, MO, MT, NC, NJ, NM, NV, OH, SC, SD, TN, TX, VA, WA, and WY.
- Product codes and lots
- Software Version 7.2.1800.3 and 7.2.1825.5