Z-0573-2011

TRABECULAR METAL" SHOULDER REAMER CASE used for implantation of the Zimmer Trabecular Metal Humeral and/or...

Zimmer GmbH / initiated 2010-09-10 / Terminated / root cause: design

Zimmer Inc. began a recall of TRABECULAR METAL" SHOULDER REAMER CASE used for implantation of the Zimmer Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Shoulder Systems... on . The reason FDA records is "The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0573-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
TRABECULAR METAL" SHOULDER REAMER CASE used for implantation of the Zimmer Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Shoulder Systems: Catalog Number 00-4312-020-00. The intramedullary reamer is used to open the mid to distal humeral canal to accept the implant.
Product code
KWT Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented
Reason for recall
The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant, delay in surgery, humeral fracture, or the need for cement in a press-fit application.
Root cause tag
design
FDA root cause
Device Design
Action
Zimmer, Inc. sent an URGENT DEVICE CORRECTION letter dated September 10, 2010 to surgeons, risk managers, and distributors. The letter identified the product, the problem, and the action to be taken. Sales Representatives were instructed to: a. provide the Dear Surgeon (Urgent Device Correction) letter and revised surgical technique directly to surgeons and risk managers at facilities using the device or b. provide a list of facilities and surgeons using the TM Humeral and TM Reverse shoulder systems to the Zimmer Recall Coordinator, and c. destroy previous versions of the surgical technique, and d. inform users of when the new surgical technique, instrument case, distal reamers, and distal pilots will be shipped to them based on communication from Zimmer, Inc. Handling of returned product: HCPs were notified to return all of the distal pilots in the scope back to Zimmer which were to be be destroyed once the recall has been completed. They will also be instructed to either return the reamers back to Zimmer or to remove these reamers from the TM Reverse shoulder systems sets and place into the Bigliani/Flatow Complete Shoulder Solution set. New surgical techniques will also be introduced and the user will be asked to remove previous versions of the surgical technique from all points of use. This removal will be verified by Zimmer Sales Representatives. For questions regarding this recall call (574) 371-8852.
Quantity in commerce
13,093 total reamers in distribution.
Distribution
Worldwide Distribution - USA, Nationwide, Switzerland and Eschbach, Germany, Switzerland, Australia, China, India, Japan, Korea, Malaysia, Singapore, Taiwan, Brazil, Venezuela, El Salvador, Colombia, Canada, and Thailand.
Product codes and lots
Catalog Number 00-4312-020-00.   All lot numbers.