Z-0307-2011
WalkMed Triton Pole Mount Infusion Pump, Catalog 300000. Indicated for delivering infusions of medications...
WalkMed Technologies, LLC / initiated 2010-06-17 / Terminated / root cause: design
Walkmed Infusion LLC began a recall of WalkMed Triton Pole Mount Infusion Pump, Catalog 300000. Indicated for delivering infusions of medications, as well as providing total parenteral nutrition... on . The reason FDA records is "Pump door may be in a near shut position, but unlatched, and the "Door Open" alarm may not sound". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0307-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- WalkMed Technologies, LLC
- Product
- WalkMed Triton Pole Mount Infusion Pump, Catalog 300000. Indicated for delivering infusions of medications, as well as providing total parenteral nutrition, and patient controlled analgesia.
- Product code
- FRN Pump, Infusion
- Reason for recall
- Pump door may be in a near shut position, but unlatched, and the "Door Open" alarm may not sound.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- All customers were notified by letter on 06/17/2010, describing the affected product and advising them of the possible malfunctioning alarm in the case a user does not follow the instructions as specified in the Triton Operation Manual. The letter also reminded all users to follow the procedures in the IFUs. The firm stated that it was best if customers re-educated all nurses on the proper door closure procedure to help prevent the potential bypass issue from occurring. Attached to the letter were two pages explaining the proper process for closing the pump door. Follow-up phone calls were made to coordinate an upgrade to all affected pumps. Questions or concerns should be directed to Ross Kurz at 720-531-4933.
- Quantity in commerce
- 1565 pumps
- Distribution
- Nationwide Distribution -- AL, CA, FL, and PA.
- Product codes and lots
- Serial numbers: 1 thru 500, TR1041 through TR2559.