Z-0350-2011

Elekta Synergy XVI

Elekta Limited / initiated 2010-09-01 / Terminated

Elekta, Inc. began a recall of Elekta Synergy XVI Intended use: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner on . The reason FDA records is "It is possible to do couch corrections twice if both Couch Move Assistant in MOSAIQ and Remote Automatic Table in XVI are configured when in SYNERGISTIQ mode". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0350-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Elekta Synergy XVI Intended use: Radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
It is possible to do couch corrections twice if both Couch Move Assistant in MOSAIQ and Remote Automatic Table in XVI are configured when in SYNERGISTIQ mode.
FDA root cause
Other
Action
Elekta sent an "Important Notice A340" dated September 1, 2010 to all customers. The notice described the product, problem and actions to be taken by the customer. The customers were instructed to make a decision if XVI or MOSAIQ does the corrective table movement; configure and ensure that the system only uses RATM or CMA for the corrective table movement, when in SYNERGISTIQ mode, and complete and return the IMPORTANT NOTICE Confirmation of Receipt form via fax at +44(0)1293 654401. If you have any questions contact your local service representative at +44(0)1293 654200.
Quantity in commerce
23 units
Distribution
Nationwide distribution: AZ, CA, DC, IL, KY, MA, MI, MS, NV, NY, OH, OR, PA, VT, and WV.
Product codes and lots
Serial Numbers 152019, 151866, 151259, 151522, 151969, 152192, 152126, 152115, 152116, 152176, 151931, 152177, 151523, 151626, 151842, 152158, 152079, 152080, 151250, 152064, 152207, 151628, and 151362