Z-0109-2011
in2it (I) A1c Test Cartridges (Catalog 281-0001EX) and in2it (II) A1c Test Cartridges (Catalog 281-0001)...
Bio-Rad Laboratories Inc. / initiated 2010-09-07 / Terminated / root cause: design
Bio-Rad Laboratories Inc began a recall of in2it (I) A1c Test Cartridges (Catalog 281-0001EX) and in2it (II) A1c Test Cartridges (Catalog 281-0001), manufactured by Bio-Rad Laboratories Deeside Ltd... on . The reason FDA records is "Test Cartridges may leak and have the potential to generate error messages and biased inaccurate results". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0109-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bio-Rad Laboratories Inc.
- Product
- in2it (I) A1c Test Cartridges (Catalog 281-0001EX) and in2it (II) A1c Test Cartridges (Catalog 281-0001), manufactured by Bio-Rad Laboratories Deeside Ltd., Flintshire, UK, distributed by Bio-Rad Laboratories, Hercules, CA. The in2it (I) is for Physician use, the (II) is for prescription home use. The products are medical devices used in affinity chromatography method, intended for the in-vitro quantitative determination of A1c (HbA1C) in capillary blood taken from a finger prick. The test is indicated for monitoring the time averaged blood glucose levels of known diabetics, for professional use as an indicator of overall glycemic control.
- Product code
- LCP Assay, Glycosylated Hemoglobin
- Reason for recall
- Test Cartridges may leak and have the potential to generate error messages and biased inaccurate results.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- BIO-RAD Laboratories Deeside sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated September 7, 2010, via registered mail to all subsidiaries/customers. The letter described the product, problem and actions to be taken by the customer. The customers were instructed to stop using and discard the test cartridges from the listed lots; complete and return the attached CUSTOMER MEDICAL DEVICE CORRECTION RESPONSE FORM via fax +44 1244 833401; record the number of affected kits of product, confirm that they have discarded the cartridges and record the method of disposal. If you have any questions, contact +44 1244 288888.
- Quantity in commerce
- 4056
- Distribution
- Worldwide distribution: USA and countries including: Australia, New Zealand, Taiwan, France, the UK, Singapore, Germany, Italy, Thailand, China, Korea, Sweden, Sri Lanka, Maldives, Canada and the Philippines.
- Product codes and lots
- Product code LCP,expiration dates between 2011-07-12 and 2011-07-26. Kit lot numbers (I):072T97, 072T98, 072T100, (II): 074P75, 074P73, Device lot numbers 131Q203 38864, 131Q204 39008, 131Q206 39088