Z-2607-2011

Eon Mini Neurostimulation (IPG) System Model 3788 System is indicated as an aid in the management of chronic...

Abbott Medical / initiated 2011-05-24 / Terminated / root cause: manufacturing process

Advanced Neuromodulation Systems Inc. began a recall of Eon Mini Neurostimulation (IPG) System Model 3788 System is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs... on . The reason FDA records is "Defective battery was identified as cause for reports of inability to communicate or recharge the product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2607-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Eon Mini Neurostimulation (IPG) System Model 3788 System is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
Defective battery was identified as cause for reports of inability to communicate or recharge the product.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
St. Jude Medical sent an "IMPORTANT MEDICAL DEVICE RECALL" letter dated May 24, 2011 to all affected customers. The letter describes the product, problem and actions to be taken by the customers. St. Jude Medical recommends no action be taken if the product serial numbers do not match the numbers listed in the letter. For product remaining in unimplanted inventory, customers are instructed to contact their SJM representative so that product may be returned and replacements be provided. For implanted product that matches the serial numbers, the Medical Advisory Board recommend not to unnecessarily explant the devices if they are functioning as intended. Replacement devices will be provided at no additional cost to the customer. A Product Reconciliation Form was attached to the letter for customers to complete and return any unused product to the St. Jude Medical Neuromodulation Division at 6901 Preston Road., Plano, TX 75024. Questions regarding this recall are directed to the customer's St. Jude Medical Neuromodulation Division Representative.
Quantity in commerce
17,141
Distribution
Worldwide Distribution-- USA (nationwide) including PUERTO RICO and countries of AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DENMARK, FINLAND,FRANCE, GERMANY, GREECE, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, MEXICO, NETHERLANDS, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TURKEY, and UNITED KINGDOM.
Product codes and lots
Serial numbers 701029 through 13766511. Serial numbers containing supplier battery lot numbers 285 and below.