Volcano s5/s5i imaging system Family Software version 3.2, Part Numbers 435-0602.01 to 05, including...
Volcano Corporation began a recall of Volcano s5/s5i imaging system Family Software version 3.2, Part Numbers 435-0602.01 to 05, including ChromaFlo and Virtual Histology, manufactured by Volcano... on . The reason FDA records is "Software defect results in improper image brightness, which may result in introduction of artifacts, which may cause misclassification of results and affect classification of tissue types. This may result in...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0168-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Volcano Corporation
- Product
- Volcano s5/s5i imaging system Family Software version 3.2, Part Numbers 435-0602.01 to 05, including ChromaFlo and Virtual Histology, manufactured by Volcano Corporation of Rancho Cordova, CA.
- Product code
- IYO System, Imaging, Pulsed Echo, Ultrasonic
- Reason for recall
- Software defect results in improper image brightness, which may result in introduction of artifacts, which may cause misclassification of results and affect classification of tissue types. This may result in misdiagnosis and inappropriate treatment.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Volcano issued a Voluntary Recall of Volcano s5 Family Software version 3.2 letter dated June 30, 2010, identifying the software defect and actions to be taken by Volcano. Field Service Engineers will visit each site and reinstall the previous software version, which will resolve the issue. Customers can contact Volcano at 916 281-2790.
- Quantity in commerce
- 79 software kits
- Distribution
- Worldwide Distribution: USA, and the countries of France, Germany, Japan, and the UK.
- Product codes and lots
- Software version 3.2, Part Numbers 435-0602.01 to 05,