Z-0331-2011

Nail Holding Screw; 1 unit per package Stryker Trauma GmbH; Distributed by Howmedica Osteonics Corp., 325...

Howmedica Osteonics Corp / initiated 2010-08-30 / Terminated

Stryker Howmedica Osteonics Corp. began a recall of Nail Holding Screw; 1 unit per package Stryker Trauma GmbH; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 The nail holding... on . The reason FDA records is "Variations in the outer diameter of the shaft of the External Compression Device and the inner diameter of the Nail Holding Screw may make assembly of these 2 instruments not possible". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0331-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Nail Holding Screw; 1 unit per package Stryker Trauma GmbH; Distributed by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 The nail holding Screw is used to assemble the nail to the target device. The Nail holding Screw is further the mating part (inner thread) of the External Compression Device as the instruments needs to be threaded into the Nail holding Screw in order to achieve Compression at the fracture side.
Product code
HSB Rod, Fixation, Intramedullary And Accessories
Reason for recall
Variations in the outer diameter of the shaft of the External Compression Device and the inner diameter of the Nail Holding Screw may make assembly of these 2 instruments not possible.
FDA root cause
Other
Action
Stryker sent Recall notification letters by Federal Express on September 9, 2010 to all Stryker branches/agencies and Risk Managers, Chief of Orthopaedics of affected hospitals and Surgeons on September 10, 2010, that used the product. The letter identified the product, the problem, and the action that customers should take. Customers were instructed to: Examine their inventory and hospital locations to identify the affected product and return it to their branch or agency warehouse for reconciliation. Reconcile all product on the attached Product Recall Acknowledgment Form and fax a copy to (201) 831-6069 within 5 days of receipt of notice. A Stryker representative would contact the customer with details for return of the product. For any questions regarding this recall call (201) 972-2100.
Quantity in commerce
22
Distribution
Nationwide Distribution including AZ, CA, CO, FL, GA, IL, KY, MD, OH, OK, and TX
Product codes and lots
Catalog Number: 1806-1602; Lot codes: K101924, K240627.