Z-0309-2011
3.5mm LCP Periarticular Proximal Humerus Plates Indicated for fractures and fracture dislocations...
Synthes GmbH / initiated 2010-09-16 / Terminated / root cause: manufacturing process
Synthes USA (HQ), Inc. began a recall of 3.5mm LCP Periarticular Proximal Humerus Plates Indicated for fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus... on . The reason FDA records is "screw can pass through plate hole". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0309-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- 3.5mm LCP Periarticular Proximal Humerus Plates Indicated for fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus, particularly for patients with osteopenic bone
- Product code
- HRS Plate, Fixation, Bone
- Reason for recall
- screw can pass through plate hole
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Synthes (USA) notified all Synthes Trauma Hospital accounts of record of this recall via an Urgent Medical Device Recall letter dated September 27, 2010. The notification included instructions for hospital accounts to immediately cease use of the product, examine inventory, remove the 3.5 mm LCP Periarticular Proximal Humerus Plates from their shelves and complete and return the attached Verification Section form via mail: Credit/Returns, Synthes, 1101 Synthes Avenue, Monument, CO 80132; fax: 610-215-9005 or Scan/email: [email protected]. If they have any of the identified devices, they were to call Synthes at 1-800-479-6329 to obtain Authorization number prior to completing Verification form. If you have any questions, please call 1-800-620-7025 x5403 or contact your Synthes Trauma Sales Consultant.
- Quantity in commerce
- 2849 units
- Distribution
- Worldwide distribution: USA and countries including: Bermuda, Korea, Italy, Japan, Georgia, Qatar and Turkey
- Product codes and lots
- Catalog/model numbers: 02.123.020, 02.123.020S, 02.123..021, 02.123.021S, 02.123.022, 02.123.022S, 02.123.023, 02.123.023S, 02.123.024, 02.123.024S, 02.123.025, 02.123.025S, 02.123.026, 02.123.026S, 02.123.027, 02.123.027S, 02.123.028, 02.123.028S, 02.123.029, 02.123.029S, 02.123.030, 02.123.030S, 02.123.031, 02.123.031S, 02.123.032, 02.123.032S, 02.123.033, 02.123.033S, 02.123.040, 02.123.040S, 02.123.041, 02.123.041S, 02.123.042, 02.123.042S, 02.123.043, and 02.123.043S. Lot numbers range form 6371761 through 6443098.