SoftMic Versions 4.0.4.7, 4.5.1.3. Version 4.0.4.7 Manufactured 02/03/2010. Version 4.5.1.3 Manufactured...
SCC Soft Computer began a recall of SoftMic Versions 4.0.4.7, 4.5.1.3. Version 4.0.4.7 Manufactured 02/03/2010. Version 4.5.1.3 Manufactured 03/18/2010. SCC Soft Computer 5400 Tech Data Drive... on . The reason FDA records is "Softmic system may delay or omit reporting of clinically significant results including organisms and drug sensitivities. Incomplete culture results may appear complete. SCC Soft Computer is issuing a Correction for...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0088-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SCC Soft Computer
- Product
- SoftMic Versions 4.0.4.7, 4.5.1.3. Version 4.0.4.7 Manufactured 02/03/2010. Version 4.5.1.3 Manufactured 03/18/2010. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760.
- Product code
- LNX
- Reason for recall
- Softmic system may delay or omit reporting of clinically significant results including organisms and drug sensitivities. Incomplete culture results may appear complete. SCC Soft Computer is issuing a Correction for SoftMic Versions 4.0.4.7-4.5.1.3 software due to this potential hazard.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- SCC Soft Computer issued Risk-to-Health notifications to clients on July 19, 2010 via proprietary communication software. Corrective actions will be via programming correction via hot fix (software modification) or such so that Global variable is not used. Clients are asked to acknowledge receipt of this task and update the task with permission to load the required correction.
- Quantity in commerce
- 35
- Distribution
- Worldwide Distribution: USA, including the states of CA, CT, DE, FL, IA, KY, MA, MA, MI, MN, NJ, NY, OH, PA, TX, VA, and WI, and the country of Canada.
- Product codes and lots
- Versions 4.0.4.7, 4.5.1.3.