SoftLab/Mic ASCII Add-on Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16. SCC Soft Computer 5400 Tech Data Drive...
SCC Soft Computer began a recall of SoftLab/Mic ASCII Add-on Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Version 4.0.1.14 Manufactured... on . The reason FDA records is "When a test is being verified, the wrong (expired) ranges may be associated to that test result, causing an incorrect flag to be posted for that test result". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0344-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SCC Soft Computer
- Product
- SoftLab/Mic ASCII Add-on Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. Version 4.0.1.14 Manufactured & Distributed 12/29/2006. Version 4.0.1.15 Manufactured & Distributed 05/02/2007. Version 4.0.1.16 Manufactured & Distributed 08/22/2008. Used in a medical research or clinical laboratory setting by knowledgeable, trained and experienced personnel.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- When a test is being verified, the wrong (expired) ranges may be associated to that test result, causing an incorrect flag to be posted for that test result.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Clients were notified via the proprietary communication software on 02/16/2010. The Notification Verbiage identified the affected product and informed clients of the potential hazard, root cause, corrective action, and requested action. Clients were instructed to acknowledge receipt of the notification. If the client wishes to receive the correction, they should document a request to load the software.
- Quantity in commerce
- 31
- Distribution
- Nationwide Distribution - US and Puerto Rico.
- Product codes and lots
- Versions 4.0.1.14, 4.0.1.15, and 4.0.1.16.