Microtek Medical, Inc., Microscope Drape for Zeiss Standard Model/Catalog Number: 70890655 Product...
Microtek Medical, Inc. began a recall of Microtek Medical, Inc., Microscope Drape for Zeiss Standard Model/Catalog Number: 70890655 Product Description: The Microtek Medical, Inc., Microscope Drape... on . The reason FDA records is "This product is being recalled due the product is miss-labeled as being sterilized by EO processing; however, the product was processed using Gamma sterilization". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0592-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Microtek Medical LLC
- Product
- Microtek Medical, Inc., Microscope Drape for Zeiss Standard Model/Catalog Number: 70890655 Product Description: The Microtek Medical, Inc., Microscope Drape for Zeiss Standard is intended is to protect surgical microscopes from contamination during various procedures. Labeling for the product is as follows: (1.) INNER CARTON LABEL: Distributed by; MICROTEX, MICROSCOPE DRAPE FOR ZEISS STANDARD, 20 x 66 (51cm x 168 cm), REF 70890655, QTY 12, LOT DA102032, Microtek Medical, Inc., Columbus, MS 39702, www.microtekmed.com, CE0044, EC, REP, Microtek Medical B.V. Zutphen, The Netherlands, Manufactured in the Dominican Republic of U.S. Materials, STERILE, EO, 2015-07, (01)30748426064206, (17) 150700(10) DA102032, INNER, Rev. C. (2.) PRODUCT LABEL: Distributed by; MICROTEX, MICROSCOPE DRAPE FOR ZEISS STANDARD, 20 x 66 (51cm x 168 cm), REF 70890655, QTY 1, LOT DA102032, Microtek Medical, Inc., Columbus, MS 39702, www.microtekmed.com, CE0044, EC, REP, Microtek Medical B.V. Zutphen, The Netherlands, Manufactured in the Dominican Republic of U.S. Materials, STERILE, EO, 2015-07, (01) 00748426064205, (17) 150700(10) DA102032, EACH, Rev. D. (3.) CARTON LABEL: Distributed by; MICROTEX, MICROSCOPE DRAPE FOR ZEISS STANDARD, 20 x 66 (51cm x 168 cm), REF 70890655, QTY 12/CS, LOT DA102032, Microtek Medical, Inc., Columbus, MS 39702, www.microtekmed.com, CE0044, EC, REP, Microtek Medical B.V. Zutphen, The Netherlands, Manufactured in the Dominican Republic of U.S. Materials, STERILE, EO, 2015-07, (01)50748426084200, (17) 150700(10) DA102032, INNER, Rev. D.
- Product code
- KKX Drape, Surgical, Antimicrobial
- Reason for recall
- This product is being recalled due the product is miss-labeled as being sterilized by EO processing; however, the product was processed using Gamma sterilization.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- On 08/23/10 Microtek Medical, Inc., Jacksonville, Florida notified all of their customers via a telephone call and requested immediate return of the product . The firm followed up with each of the customers by sending out an Urgent: Medical Device Recall letter (written notice), dated September 02,2010, to each on 9/8/10. The letter identified the affected product, stated the reason for the recall, and asked that the product be returned immediately. Customers are to complete and return the enclosed response form. Anyone needing further assistance is asked to contact Elizabeth Jolly, V.P. Legal at 1-888-844-0988.
- Quantity in commerce
- 5 Cases, 12 drapes per case, 60 Drapes total distrubuted
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot Number: DA102032 Expiration Date: 2015-07