Maquet Heartstring II Proximal Seal System (HS-1045) manufactured by Maquet Cardiovascular, Wayne, NJ
Maquet Cardiovascular, LLC began a recall of Maquet Heartstring II Proximal Seal System (HS-1045) manufactured by Maquet Cardiovascular, Wayne, NJ Intended use: medical device for use by cardiac surgeons... on . The reason FDA records is "Deployment tube may detach during use due to insufficient adhesive, which may result in product failure to operate". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0940-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular LLC
- Product
- Maquet Heartstring II Proximal Seal System (HS-1045) manufactured by Maquet Cardiovascular, Wayne, NJ Intended use: medical device for use by cardiac surgeons during CABG procedures to maintain hemostasis and to facilitate the completion of a proximal anastomosis without application of an aortic clamp.
- Product code
- DXC Clamp, Vascular
- Reason for recall
- Deployment tube may detach during use due to insufficient adhesive, which may result in product failure to operate.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- The firm, MAQUET Getinge Group, sent an "URGENT DEVICE REMOVAL IMMEDIATE ACTION REQUIRED" letter dated May 5, 2010, to all customers via Federal Express. The letter described the product, problem and actions to be taken by the customer. The customers were instructed to immediately examine their stocks to determine if they have any of the affected product, if found, discontinue dispensing (distributing) the lot and complete the Field Action Response Form; immediately contact accounts if they acted as a distributor, and acknowledge receipt of this notification whether or not they have any affected products by completing the Field Action Response Form via fax to (U.S.): 1-888-899-2874. If you have any additional questions, please contact your local MAQUET Cardiovascular sales representative or our Customer Service at 1-888-880-2874.
- Quantity in commerce
- 1630
- Distribution
- Worldwide distribution: USA and countries including: Japan, Mexico, Germany, Switzerland, Israel
- Product codes and lots
- Lot numbers 12895788, 12946718, 25002866, 25004647, 13029141, 25005045, 25005622