LeMaitre Vascular AnastoClip GC Vessel Closure System-Large Size Models: 4007-03 Large size 3 Pack 4008-03...
Lemaitre Vascular, Inc. began a recall of LeMaitre Vascular AnastoClip GC Vessel Closure System-Large Size Models: 4007-03 Large size 3 Pack 4008-03 Large size 10 Pack 4007-07 Large size 1 Pack... on . The reason FDA records is "Mislabeled size on external packaging: The contents section listed on the Tyvek lid labels of the Medium and Large sizes incorrectly identified the size of the clip as XL, while the device inside was either a Medium or...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0165-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LeMaitre Vascular Inc.
- Product
- LeMaitre Vascular AnastoClip GC Vessel Closure System-Large Size Models: 4007-03 Large size 3 Pack 4008-03 Large size 10 Pack 4007-07 Large size 1 Pack Manufactured by LeMatire Vascular Inc, Burlington, MA 01803
- Product code
- HBT Applier, Hemostatic Clip
- Reason for recall
- Mislabeled size on external packaging: The contents section listed on the Tyvek lid labels of the Medium and Large sizes incorrectly identified the size of the clip as XL, while the device inside was either a Medium or Large size.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- LeMaitre Vascular issued an Urgent AnastoClip GC Device Field Safety Notice letter dated Sepember 28, 2010 identifying the affected devices and actions to be taken by customers. The letter contains a form that is requested to be returned to LeMaitre Vascular as record of notification and reconciliation. The letter provides an option for the notified hospital to return the affected product for replacement if unused or if they simply wish to have it corrected. Customers can contact LeMaitre Customer Service for replacement at 800 628-9470.
- Quantity in commerce
- 108 units
- Distribution
- Nationwide Distribution: to the states of CA, FL, GA, IL, IN, LA, MD, MS, NJ, OK, OR, and TX.
- Product codes and lots
- Lot Number Expiration Date AGC1015 2012-10 AGC1016 2012-10 AGC1017 2012-10 AGC1018 2012-11 AGC1019 2012-11 AGC1022 2012-11 AGC1023 2012-11 AGC1024 2012-12 AGC1025 2013-01 AGC1027 2013-01 AGC1028 2013-01