Z-1021-2011

Vital Connect 4.0 & 4.1, radiological image processing software A Medical diagnostic software system intended...

Canon Medical Informatics, Inc. / initiated 2010-08-23 / Terminated / root cause: software

Vital Images, Inc. began a recall of Vital Connect 4.0 & 4.1, radiological image processing software A Medical diagnostic software system intended to process, analyze, review, and distribute... on . The reason FDA records is "There is a potential alignment error of fused PET/CT images in the Vitrea Core component of Vitrea Enterprise Suite 1.2 and 1.3. The error occurs during panning of the fused PET/CT images. Because the PET data has a...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1021-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Informatics, Inc.
Product
Vital Connect 4.0 & 4.1, radiological image processing software A Medical diagnostic software system intended to process, analyze, review, and distribute multi-dimensional digital images acquired from a variety of imaging devices including: CT MR, CR/DR/DX, SC, US, NM, PET, XA and RF, etc. Vital connect has the following intended use: Vessel Probe is intended for viewing the anatomy and pathology of a patient's coronary arteries. Clinicians select any artery to view the following anatomical references: the highlighted vessel in 3D, two rotate-able curved MPR vessel views displayed at angles orthogonal to each other, and cross sections of the vessel.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
There is a potential alignment error of fused PET/CT images in the Vitrea Core component of Vitrea Enterprise Suite 1.2 and 1.3. The error occurs during panning of the fused PET/CT images. Because the PET data has a larger voxel size than the CT data, it is possible to pan the CT image in more frequent, smaller steps than the PET image, so that the fused PET image does not move until the CT data
Root cause tag
software
FDA root cause
Software design
Action
The firm, Vital Images, Inc,., sent a "Urgent Software Alert" letter dated August 23, 2010 to all customers. The letter described the product, problem and actions to be taken by the customers. The customers were instructed to not use the PET/CT fusion functionality until the software updates are provided and to re-examine any past results. If you have any questions or comments, contact Customer Support at [email protected] or at 1-800-208-3005.
Quantity in commerce
178
Distribution
Worldwide Distribution: USA including states of: AK, AZ, CA, CO, CT, FL, GA, IL, IN, IA, KY, LA, ME, MD, MA, MI, MN, MS, MO, NH, NJ, NM, NY, NC, ND, OH, PA, SC, TN, TX VA, WV, WI, and WY; and countries of: AUSTRALIA, HONG KONG, UNITED KINGDOM, NETHERLANDS, FRANCE, BELGIUM, CANADA, CHINA, AUSTRALIA, SWITZERLAND, and ITALY.AUSTRALIA, HONG KONG, UNITED KINGDOM, NETHERLANDS, FRANCE, BELGIUM, CANADA, CHINA, AUSTRALIA, SWITZERLAND, ITALY.
Product codes and lots
Any VitalConnect version, or the Vitrea Core component of Vitrea Enterprise Suite (VES) 1.2 adn VES 1.3.