Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0361-2011

ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Model # ATS45. Manfuactured by Ethicon...

Ethicon Endo-Surgery Inc / initiated 2009-01-13 / Terminated

Ethicon Endo-Surgery Inc began a recall of ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Model # ATS45. Manfuactured by Ethicon Endo-Survery, LLC; 475 Calle C; Guaynabo PR 00969 on . The reason FDA records is "The mechanism that connects the articulation joint to the device shaft may cause the jaws of the cutters to remain closed and clamped down on tissue after the device is fired. If the jaws of the device remain clamped...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0361-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon Endo-Surgery Inc
Product
ENDOPATH ETS-FLEX45 Articulating Endoscopic Linear Cutters, Model # ATS45. Manfuactured by Ethicon Endo-Survery, LLC; 475 Calle C; Guaynabo PR 00969.
Product code
GDW Staple, Implantable
Reason for recall
The mechanism that connects the articulation joint to the device shaft may cause the jaws of the cutters to remain closed and clamped down on tissue after the device is fired. If the jaws of the device remain clamped down on the tissue, there is a risk the surgery will need to be modified to remove the device. This may lead to an increase in procedure time, tissue manipulation, and a change in t
FDA root cause
Other
Action
Ethicon Endo-Surgery issued "Urgent Device Recall Event 1952" notification letters dated January 21, 2009 to their customers. They instructed their custormers to immediately discontinue use of the product. They are to complete and fax back the Business Reply Form. Affected product is to be returned with a copy of the completed Business Reply Form to Stericycle. Replacement product should be received within 30-60 days after return. Customers can contact Stericycle concerning this recall at 1-866-861-6921, reference Event 1952.
Quantity in commerce
4,113 UNITS
Distribution
Worldwide Distribution: USA only, in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI. IA. ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.
Product codes and lots
Lot #'s: E4MD1T, E4MD5W, E4MD3Y, E4ME7J, E4MD4Z, E4ME7U, E4MD50 & E4MF0A