Z-0988-2011

Southmedic/Personna Safety Scalpel with Personna Plus Blades, Reorder No. 73-0610, Size #10, General Surgical...

Southmedic Inc. / initiated 2010-09-03 / Terminated / root cause: packaging

Southmedic, Inc. began a recall of Southmedic/Personna Safety Scalpel with Personna Plus Blades, Reorder No. 73-0610, Size #10, General Surgical Use, Sterile, unless package is damaged or... on . The reason FDA records is "Scalpels labeled as sterile have the potential to be unsealed and therefore not sterile". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0988-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Southmedic Inc.
Product
Southmedic/Personna Safety Scalpel with Personna Plus Blades, Reorder No. 73-0610, Size #10, General Surgical Use, Sterile, unless package is damaged or opened., Southmedic Inc., 50 Alliance Blvd., Barrie, ON., Canada L4M 5K3 www.southmedic.com, Made in Canada.
Product code
GDX Scalpel, One-Piece
Reason for recall
Scalpels labeled as sterile have the potential to be unsealed and therefore not sterile.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
On 9/3/2010, the firm notified consignees about the recall by facsimile letter. The letter instructed consignees to return scalpels for credit or replacement. The firm enlarged the scope of the recall on 11/15/2010. An additional 927 cases were recalled. This includes 11 additional lots. The firm sent out URGENT MEDICAL DEVICE RECALL letters dated 11/15/2010 to its direct accounts. The letter identified that affected product and the reason for the recall. The consignees were instructed to check their inventory, complete and return the attached Voluntary Recall Notification Confirmation Form, return the affected product to the manufacturer. Product return question should be directed towards Marc Pauze at 800-463-7146 ext 299 and regulatory questions should be directed towards Tish Anger at 800-463-7146 ext 342.
Quantity in commerce
1330 all products
Distribution
Worldwide Distribution -- USA, including states of TX, MO, NJ, LA, IL, NY, KY, PA, VA, WA, CA, MA, MD, IN, NV, TN, NC, AZ, IA, WI, UT, NE, GA, FL, MN, ME, OH, SD, and KS and countries of Australia and Canada.
Product codes and lots
Lot numbers: 070610, 82010, 82610, and 91710.