Z-1668-2011

Abbott brand CELL-DYN Emerald Diluent Reagent, 10 mL, List Number: 09H48-02, Product is manufactured and...

Abbott Medical / initiated 2010-09-30 / Terminated / root cause: labeling and UDI

Abbott Laboratories began a recall of Abbott brand CELL-DYN Emerald Diluent Reagent, 10 mL, List Number: 09H48-02, Product is manufactured and distributed by Abbott Diagnostics Division, Santa... on . The reason FDA records is "The product bar code label, when scanned, indicates that the volume is 10 mL rather than the correct volume of 10,000 mL. Analytical results are not impacted, although, there is a potential for delay in generating...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1668-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott brand CELL-DYN Emerald Diluent Reagent, 10 mL, List Number: 09H48-02, Product is manufactured and distributed by Abbott Diagnostics Division, Santa Clara. CA Designed for use on the CELL-DYN Emerald System, which is an automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories.
Product code
GKZ Counter, Differential Cell
Reason for recall
The product bar code label, when scanned, indicates that the volume is 10 mL rather than the correct volume of 10,000 mL. Analytical results are not impacted, although, there is a potential for delay in generating results.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Product Recall notification letters were sent on 9/30/2010. The letter identified the affected product, an explanation of the problem, patient impact, and necessary actions. Consignees are to contact their local hematology customer support representative to arrange for replacement of the CELL-DYN Emerald Diluent Lot. Once the replacement lot has been received, customers should destroy the affected lot according to their laboratory procedures. A copy of the letter should be retained for their records and the Customer Reply Form should be completed and returned via fax or e-mail. US customers should contact Customer Support at 1-877-4ABBOTT, if they have questions.
Quantity in commerce
322 units
Distribution
Worldwide Distribution -- USA, including states of PA, NY, CA, NC, IL, TX, KS, UT, VA, FL, AR, TN, MN, NE, MI, and DE and the country of Germany.
Product codes and lots
Lot Number: 4350, Expiration Date: June 30, 2012