BD Vacutainer Push Button collection Set; Non-Pyrogenic, Sterile, Rx only; 23 G x 3/4 x 12" (0.6 x 19 mm x...
Becton Dickinson & Company began a recall of BD Vacutainer Push Button collection Set; Non-Pyrogenic, Sterile, Rx only; 23 G x 3/4 x 12" (0.6 x 19 mm x 305 mm) Becton Dickinson & Co., Franklin Lakes, NJ... on . The reason FDA records is "There is the possibility for certain lots of the BD Vacutainer Push Button Blood Collection sets to have separation of the cannula from the hub". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0816-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- BD Vacutainer Push Button collection Set; Non-Pyrogenic, Sterile, Rx only; 23 G x 3/4 x 12" (0.6 x 19 mm x 305 mm) Becton Dickinson & Co., Franklin Lakes, NJ, USA Intended use: Blood collection
- Product code
- FMI Needle, Hypodermic, Single Lumen
- Reason for recall
- There is the possibility for certain lots of the BD Vacutainer Push Button Blood Collection sets to have separation of the cannula from the hub.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- BD Diagnostics sent "RECALL NOTIFICATION" separate letters October 1 and 5, 2010 and follow up letter dated October 8, 2010, as well as a letter dated December 1, 2010 to all distributors and customers commencing October 5, 2010 by federal express. The letters describe the product, the problem and the action to be taken by the customers. The customers were instructed to immediately examine their inventory for specific catalog and lot numbers, and return product from the lots following the instructions on the enclosed packing slip (BD will send replacement upon receipt of returned product). If you have any questions on returning and replacing product, please contact BD logistics at 1-866-380-0046. If you have any further detail questions regarding this recall, please contact your BD Sales Consultant or BD Technical Services via email at www.BD.com/Vacutainer/Contact or via telephone at 1-800-631-0174.
- Quantity in commerce
- 4,223,200
- Distribution
- Nationwide Distribution
- Product codes and lots
- Catalog Number 367342; Lot numbers: 0145457, 0152300, 0153981, 0153982, 0158081, 0158082, 0158084, 0160664, 0160666, all exp 5/31/2012; Lot numbers 0165202, 0165205, 0165207, 0165218, 0165226, 0165227, 0165232, 0165234, 0176872, 0179208, 0179209, 0179212, 0180023, 0180026, 0180027, 0180028, 0180040, 0180042, all exp 6/30/2012; Lot numbers 0196209, 0196210, 0201399, 0201400, 0201402, 0208593, 0211641, all exp 7/31/2012.