Captus 3000 Thyroid Uptake System with Portable Stand option. Capintec item number 5430-2130, serial numbers...

Mirion Technologies (Capintec), Inc. / initiated 2010-09-20 / Terminated / root cause: manufacturing process

Capintec Inc began a recall of Captus 3000 Thyroid Uptake System with Portable Stand option. Capintec item number 5430-2130, serial numbers: 91001, 91003 through 91010. Manufactured by... on . The reason FDA records is "The spring arm on portable stand may have been over tightened during assembly, which may result in failure. If the spring arm fails, it will not balance the collimator weight, and the arm will drop to the lowest position". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0167-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Mirion Technologies (Capintec), Inc.
Product
Captus 3000 Thyroid Uptake System with Portable Stand option. Capintec item number 5430-2130, serial numbers: 91001, 91003 through 91010. Manufactured by Capintec, Inc, Pittsburgh, PA.
Product code
KPT Calibrator, Dose, Radionuclide
Reason for recall
The spring arm on portable stand may have been over tightened during assembly, which may result in failure. If the spring arm fails, it will not balance the collimator weight, and the arm will drop to the lowest position.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Capintec issued an Alert letter to all customers identifying the affected device, informing them of the problem and what actions are to be taken by customers. Capintec will send a replacement arm for the system, which is to be installed by the customer using included instructions. The firm included a correction acknowledgement form to be completed and returned by the customer. Customers can contact Capintec concerning this recall at 201 825-9500.
Quantity in commerce
9 units
Distribution
Worldwide Distribution: USA in the states of NC and TN, and in the countries of Pakistan, Paraguay, Portugal, Saudi Arabia, and Spain.
Product codes and lots
Capintec item number 5430-2130, serial numbers: 91001, 91003 through 91010