Z-1592-2011

Straumann coDiagnostiX, dental implant planning system. Software Version 7.0 including updates 7.0.2 and...

Straumann USA, LLC / initiated 2010-09-30 / Terminated / root cause: software

Straumann Manufacturing Inc. began a recall of Straumann coDiagnostiX, dental implant planning system. Software Version 7.0 including updates 7.0.2 and 7.0.2.10 on . The reason FDA records is "Software: Surgical Error may occur when the implant length is changed after positioning the surgical sleeve". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1592-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Straumann USA, LLC
Product
Straumann coDiagnostiX, dental implant planning system. Software Version 7.0 including updates 7.0.2 and 7.0.2.10
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Software: Surgical Error may occur when the implant length is changed after positioning the surgical sleeve.
Root cause tag
software
FDA root cause
Software design
Action
Straumann notified accounts of this correction by letter dated September 30, 2010 and delivered via Federal Express Delivery Service. The upgrade is available on-line by logging into the coDiagnostiX system and enabling the automatic upgrade to occur.
Quantity in commerce
20 systems
Distribution
Nationwide
Product codes and lots
Article Number Description Version 019.0333 Offline Upgrade CDX Station 7 EN (6+) 7.0.2.10 019.0334 Offline Upgrade CDX PPP 7 EN 7.0.2.10 019.0343 coDiagnostiX Station V 7.0 EN 7.0.2.10 019.0344 coDiagnostiX PPP V 7.0 EN 7.0.2.10 019.0345 coDiagnostiX coDoctor V 7.0 EN 7.0.2.10