Millar Disposable Angiographic Catheter (Left Heart), manufactured for Millar Instruments, Inc 6001-A Gulf...
Millar Instruments, Inc began a recall of Millar Disposable Angiographic Catheter (Left Heart), manufactured for Millar Instruments, Inc 6001-A Gulf Freeway, Houston, Texas 77023 on . The reason FDA records is "The sterile packaging was found to be compromised when tested after shipping conditions. This may result in infection or transmission of disease. A test for transportation of sterile package was conducted and resulted...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0401-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Millar LLC.
- Product
- Millar Disposable Angiographic Catheter (Left Heart), manufactured for Millar Instruments, Inc 6001-A Gulf Freeway, Houston, Texas 77023
- Product code
- DQO Catheter, Intravascular, Diagnostic
- Reason for recall
- The sterile packaging was found to be compromised when tested after shipping conditions. This may result in infection or transmission of disease. A test for transportation of sterile package was conducted and resulted in several samples failing.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Millar issued a Recall Report letter on October 14, 2010 to consignees identifying the affected product, sterility issue, and actions to be taken. Customers are to return all unused catheters to Millar for replacement. Customers can contact Millar about this recall at 832-667-7131.
- Quantity in commerce
- 12 Units
- Distribution
- Worldwide Distribution: USA, including the states of MA, OH, PA, and D.C., and the countries of Canada and Germany.
- Product codes and lots
- Catalog-Models: SPC-454D, SPC-454E, SPC-454F, SPC-464D. Product Code/Part No.: 800-4219-1, 800-4219-2, 800-4219-3, 802-1029X. Lot No.: L057674, L064963, L057687, L062222, L061409, L057687, L065039, L058763. S/N: 407126, 407128, 493801, 391962, 466255, 466256, 435426, 391965, 391968, 493756, 425284, 425285.