Z-0903-2011
PROCEED Surgical Mesh; Product Code PCDG1; :Lot BMG618; Ethicon, Inc. a Johnson & Johnson Company. Used for...
Ethicon, Inc / initiated 2010-10-18 / Open, Classified / root cause: manufacturing process
Ethicon, Inc. began a recall of PROCEED Surgical Mesh; Product Code PCDG1; :Lot BMG618; Ethicon, Inc. a Johnson & Johnson Company. Used for the repair of hernias and other fascial... on . The reason FDA records is "There is a potential for delamination in one lot of Proceed surgical mesh". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0903-2011
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ethicon, Inc
- Product
- PROCEED Surgical Mesh; Product Code PCDG1; :Lot BMG618; Ethicon, Inc. a Johnson & Johnson Company. Used for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
- Product code
- FTL Mesh, Surgical, Polymeric
- Reason for recall
- There is a potential for delamination in one lot of Proceed surgical mesh.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Ethicon sent Urgent Voluntary Product Recall letters via Fed Ex on October 18, 2010. The letters described the issue with the affected product and instructed customers not to use it. They are to immediately remove the recalled product from their inventory and return it according to the instructions provided. A reply card is attached and must be filled out and returned to Stericycle whether or not the customer has product in their possession. Separate correspondence has been sent to end customers stating that if they have purchased the recalled product through a distributor, to return the product directly to Stericycle. Replacement product will be provided at no additional charge for all recalled product.
- Quantity in commerce
- 87 products
- Distribution
- Worldwide Distribution
- Product codes and lots
- Product code PCDG1, Lot number BMG618.