Z-0398-2011
SKYLight Gamma Camera Gearbox brake, a component of the SKYLight Gamma Cameras 3/8" SPECT, Model number 4535...
Philips North America LLC / initiated 2010-10-15 / Terminated
Philips Medical Systems began a recall of SKYLight Gamma Camera Gearbox brake, a component of the SKYLight Gamma Cameras 3/8" SPECT, Model number 4535 600 66661; and 5/8" SPECT, Model number 4535 602... on . The reason FDA records is "Detector arm assembly may slide to hardware limit, potentially causing impact". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0398-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- SKYLight Gamma Camera Gearbox brake, a component of the SKYLight Gamma Cameras 3/8" SPECT, Model number 4535 600 66661; and 5/8" SPECT, Model number 4535 602 20381, manufactured by Philips Medical Systems (Cleveland), Cleveland OH. Intended to produce images depicting the anatomical distributions of single photon and positron emitting radioisotopes within the human body for interpretation by medical personnel.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- Detector arm assembly may slide to hardware limit, potentially causing impact.
- FDA root cause
- Component change control
- Action
- Firm will notify consignees via Urgent - Medical Device Correction, dated Nov 3, 2010, via FedEx. The letter identified the affected product, described the problem, the hazard involved, how to identify affected products, actions to be taken, and actions planned by Philips. Customers are to contact their Field Safety Engineer immediately to schedule an inspection of their system. If customers suspect that the identified problem has occurred with their device, they should immediately stop using the system until Philips has confirmed it as safe to use. If customers need further information concerning this issue, they are to contact their local Philips representative.
- Quantity in commerce
- 24 units
- Distribution
- Worldwide Distribution -- USA and the Netherlands.
- Product codes and lots
- 3/8" SPECT, Model number 4535 600 66661, and 5/8 SPECT, Model number 4535 602 20381, all units manufactured between July 2001 to August 2003.