Philips IntelliVue Clinical Information Portfolio (ICIP), Clinical Management System Software Version D.00...
Philips Healthcare Inc. began a recall of Philips IntelliVue Clinical Information Portfolio (ICIP), Clinical Management System Software Version D.00 through D.03 on . The reason FDA records is "Software: Scheduling function of the Philips Intellivue Clinical Information Portfolio Releases D.00 through D.03. Edits to a scheduled even frequency order may result in pending interventions not being generated after...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0414-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips IntelliVue Clinical Information Portfolio (ICIP), Clinical Management System Software Version D.00 through D.03.
- Product code
- DXJ Display, Cathode-Ray Tube, Medical
- Reason for recall
- Software: Scheduling function of the Philips Intellivue Clinical Information Portfolio Releases D.00 through D.03. Edits to a scheduled even frequency order may result in pending interventions not being generated after the first 24 hours.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips Healthcare issued an Urgent Medical Device Correction letter dated October 15, 2010 to customers. Customers are asked to follow the guidelines in the "Action to be taken by Customer/User" section of the Urgent Medical Device Correction Notice until they receive their software correction. The correction will be in the form of a software upgrade and will be performed by Philips service personnel. Contact Philips Customer Care Center with questions or concerns about this correction at 1-800-722-9377
- Quantity in commerce
- 200 units
- Distribution
- Worldwide Distribution: USA, including states of CA, DC, FL, IL, IN, KY, MA, MD, ME, NJ, NY, OK, RI, TX, VA, VT, and WV; and the countries of: AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHINA, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, GERMANY, HONG KONG, IRELAND, ITALY, KENYA, LATVIA, MALAYSIA, NETHERLANDS, NORWAY, SAUDI ARABIA, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, PROVINCE OF CHINA, THAILAND, UNITED ARAB EMIRATES, and the UNITED KINGDOM.
- Product codes and lots
- Philips IntelliVue Clinical Information Portfolio Release D.XX, where XX is 00, 01, 02 or 03.