Z-0411-2011
smith & nephew TRIGEN SURESHOT Distal Targeting System, Trauma Interface, V2.0.2, Trauma Launcher V1.01.1...
Smith & Nephew, Inc. / initiated 2010-10-15 / Terminated / root cause: software
Smith & Nephew Inc began a recall of smith & nephew TRIGEN SURESHOT Distal Targeting System, Trauma Interface, V2.0.2, Trauma Launcher V1.01.1 Product is intended to be an intraoperative... on . The reason FDA records is "Targeting software has been found to provide erroneous targeting when used with the 13 mm TAN/FAN Left nails". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0411-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- smith & nephew TRIGEN SURESHOT Distal Targeting System, Trauma Interface, V2.0.2, Trauma Launcher V1.01.1 Product is intended to be an intraoperative image-guided localization system. It is a computer-assisted orthopedic surgery tool to aid the surgeon with drill positioning for screws during intramedullary nail implantation.
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- Targeting software has been found to provide erroneous targeting when used with the 13 mm TAN/FAN Left nails.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Smith & Nephew Inc. sent an URGENT MEDICAL DEVICE CORRECTION letter by FedEx on October 15, 2010, to all customers. The letter was dated October 4, 2010, and identified the product, the problem, and the action to be taken by the customers. Customers were instructed to perform a software upgrade and complete the form provided and return by fax to (901) 566-7835. For questions regarding this recall call (901) 399-5441.
- Quantity in commerce
- 445 units
- Distribution
- Worldwide Distribution - USA including AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NM, OH, PA, TN, TX, AND WA and the countries of Australia, Belgium, Canada, China, Denmark, Finland, France, Germany, Italy, Japan, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland, Turkey, and the UK.
- Product codes and lots
- All serial numbers prior to and including ACA5