smith & nephew TRIGEN SURESHOT Distal Targeting System, Trauma Interface, V2.0.2, Trauma Launcher V1.01.1...

Smith & Nephew, Inc. / initiated 2010-10-15 / Terminated / root cause: software

Smith & Nephew Inc began a recall of smith & nephew TRIGEN SURESHOT Distal Targeting System, Trauma Interface, V2.0.2, Trauma Launcher V1.01.1 Product is intended to be an intraoperative... on . The reason FDA records is "Targeting software has been found to provide erroneous targeting when used with the 13 mm TAN/FAN Left nails". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0411-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
smith & nephew TRIGEN SURESHOT Distal Targeting System, Trauma Interface, V2.0.2, Trauma Launcher V1.01.1 Product is intended to be an intraoperative image-guided localization system. It is a computer-assisted orthopedic surgery tool to aid the surgeon with drill positioning for screws during intramedullary nail implantation.
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
Targeting software has been found to provide erroneous targeting when used with the 13 mm TAN/FAN Left nails.
Root cause tag
software
FDA root cause
Other
Action
Smith & Nephew Inc. sent an URGENT MEDICAL DEVICE CORRECTION letter by FedEx on October 15, 2010, to all customers. The letter was dated October 4, 2010, and identified the product, the problem, and the action to be taken by the customers. Customers were instructed to perform a software upgrade and complete the form provided and return by fax to (901) 566-7835. For questions regarding this recall call (901) 399-5441.
Quantity in commerce
445 units
Distribution
Worldwide Distribution - USA including AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NM, OH, PA, TN, TX, AND WA and the countries of Australia, Belgium, Canada, China, Denmark, Finland, France, Germany, Italy, Japan, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland, Turkey, and the UK.
Product codes and lots
All serial numbers prior to and including ACA5