Z-0447-2011

Medtronic Navigation O-Arm Imaging System, Door Assembly (BI-700-00006) used in Product Catalog Numbers...

Medtronic Navigation, Inc / initiated 2010-10-15 / Terminated / root cause: manufacturing process

Medtronic Navigation, Inc. began a recall of Medtronic Navigation O-Arm Imaging System, Door Assembly (BI-700-00006) used in Product Catalog Numbers: 9732719,9733346,81-700-00027, 81-700-00027R, 81700-... on . The reason FDA records is "Safety Notice: Door Cable Periodic Maintenance Schedule and Cable Wear for the O-ARM Imaging System Breakage of the door cable during opening and closing may occur and cause physical harm or delay of therapy". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0447-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic Navigation O-Arm Imaging System, Door Assembly (BI-700-00006) used in Product Catalog Numbers: 9732719,9733346,81-700-00027, 81-700-00027R, 81700- 00027-100, 81-700-00027-1 OOR, 81-700-00027-120, 81-700-00027-120R, 81700-00027-230, 81-700-00027-230R, 81-700-00027GER, 81-700-00027GERR Intended use: Designed for 2D fluoroscopic and 3D imaging and is intended to be used by a physician benefits from 2D and 3D information of anatomic structures with high x-ray attenuation such as bony and metallic objects.
Product code
IZL System, X-Ray, Mobile
Reason for recall
Safety Notice: Door Cable Periodic Maintenance Schedule and Cable Wear for the O-ARM Imaging System Breakage of the door cable during opening and closing may occur and cause physical harm or delay of therapy
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Medtronic Navigation sent all consignees/customers an "Urgent Field Safety Notice" letter dated October 15, 2010. The letter described the product, problem and action to be taken by the customers. The customers were instructed to review and follow the attached information that describes the process to properly detect and proactively manage cable wear; to ensure that the door is completely open when positioning the patient into or removing them from the gantry and when placing the O-ARM Imaging System Tube Drape (BI-900-00003) onto the unit, and report any visual or audible (cracking noise) wear of the cable to Medtronic Representative ASAP. Note: A Medtronic Representative will be contacting the customers within the coming weeks to schedule a time to inspect the current condition of the cable, explain and train the customers on the recommended workflow and inspection procedure and place a permanent label on their O-ARM Imaging System to reinforce these actions. If you have any questions about this notification, please contact your local Medtronic Navigation representative or call our Technical Services Group at 800-595-9709 or 720-890-3200.
Quantity in commerce
244 units
Distribution
Worldwide distribution: USA including states of: AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, and WI; and countries including: Canada, Austria, Australia, Belgium, Brazil, Libya, South Africa, India, Japan, Singapore South Korea, Haiti, Kuwait, Saudi Arabia, Turkey, UAE, Germany, Netherlands, Great Britain, Italy, Poland, Netherlands, Spain,Sweden, and Switzerland.
Product codes and lots
All Units distributed from June 2006 through September 2010