Z-0972-2011

Bio Med Devices Disposable Pediatric Breathing Circuit, One Set Catalog Number: 4040LC 5 foot /15mm...

Bio-Med Devices, Inc. / initiated 2010-10-19 / Terminated

Biomed Devices began a recall of Bio Med Devices Disposable Pediatric Breathing Circuit, One Set Catalog Number: 4040LC 5 foot /15mm corrugated EVA hose Intended use: Breathing circuit is a... on . The reason FDA records is "Breathing circuit may disconnect between the adapter adjacent tubing cuff". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0972-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Med Devices, Inc.
Product
Bio Med Devices Disposable Pediatric Breathing Circuit, One Set Catalog Number: 4040LC 5 foot /15mm corrugated EVA hose Intended use: Breathing circuit is a pediatric disposable tubing assembly for use with Lifecare ventilators
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Breathing circuit may disconnect between the adapter adjacent tubing cuff
FDA root cause
Component change control
Action
The firm,Bio Med Devices, sent a "URGENT MEDICAL DEVICE RECALL" letter dated October 18, 2010 via fax to the customers. The letter described the product, problem and action to be taken by the customers. The customers were instructed to immediately examine their stock for the affected lots, if found, discontinue distributing and using the product and promptly return via UPS #025106 to Bio-Med Devices 61 Soundview Rd, Guilford, CT 06437, Attention: 4040LC Recall Return. The customers were also instructed to immediately contact their accounts, if the product was further distributed , advise them of the recall situation and have their outstanding product returned, and to complete and return the RETURN FAX-IMMEDIATE RESPONSE REQUIRED form via fax to 203-458-0440. Note: Bio-Med informed the customers that they will credit the return product; however, they only plan to replace the product in an emergency situation, because they have very few of the 3/8" x 15mm adapters left and are looking to discontinue the 4040LC due to the discontinuance of this product by the manufacturer. If you have any questions, please contact (203) 458-0202 x 217.
Quantity in commerce
15 cases (20/cs)
Distribution
Nationwide distribution: OH, WI.
Product codes and lots
Lot Numbers:  09072210 09081310, 09082510, 05091510