Z-0377-2011

Sorin OptiPack, Perfusion Tubing Set, Sterile EO, REF 084500501, Manufactured by Sorin Group USA, Inc. 14401...

LivaNova USA, Inc. / initiated 2010-10-07 / Terminated / root cause: packaging

Sorin Group USA, Inc. began a recall of Sorin OptiPack, Perfusion Tubing Set, Sterile EO, REF 084500501, Manufactured by Sorin Group USA, Inc. 14401 W. 65th Way Arvada, CO 80004 Indicated for use in... on . The reason FDA records is "Defective product packaging may compromise product sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0377-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Sorin OptiPack, Perfusion Tubing Set, Sterile EO, REF 084500501, Manufactured by Sorin Group USA, Inc. 14401 W. 65th Way Arvada, CO 80004 Indicated for use in surgical procedures requiring extracorporeal support for periods of up to six hours.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Defective product packaging may compromise product sterility.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Sorin Group sent an "URGENT: FIELD SAFETY NOTICE" letter dated October 7, 2010, to all customers. The letter described the product, problem and action to be taken by customer. The customers were instructed to check all affected inventory to exam for holes. Any damaged products were to be returned to Sorin. If no damage was found, product could be used. The customers were also instructed to contact Sorin Cardiopulmonary Customer Service at 1-800-650-2623 to return product and receive replacement; they were to share this information with all personnel who may use this product; and to complete and return the attached Customer Response Form via the self-addressed prepaid envelope provided or fax to (303) 467-6502 as soon as possible. For further information, contact 1-800-650-2623.
Quantity in commerce
69
Distribution
Nationwide distribution: CA, FL, IL, IN, KS, MD, MO, NE, NJ, NM, TN, TX, and VA.
Product codes and lots
Lot No. 1013100022, 1016700043, 1018200024, 1022300025