Z-1394-2011

Maquet Cardiopulmonary AG, MAQUET RPM 20: Roller Pump for Heart Lung Machine HL20

Maquet Cardiovascular Us Sales, LLC / initiated 2010-09-28 / Terminated

Maquet Cardiovascular Us Sales, Llc began a recall of Maquet Cardiopulmonary AG, MAQUET RPM 20: Roller Pump for Heart Lung Machine HL20 on . The reason FDA records is "A wrong ball bearing was installed in the pump heads of product. This may make the pump head wear out and cease function". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1394-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Maquet Cardiopulmonary AG, MAQUET RPM 20: Roller Pump for Heart Lung Machine HL20.
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
A wrong ball bearing was installed in the pump heads of product. This may make the pump head wear out and cease function.
FDA root cause
Component change control
Action
Maquet hand delivered a letter to the consignee on 10/4/2010.
Quantity in commerce
4 units
Distribution
One hospital in New York City
Product codes and lots
Device Part Number (Model Number): 703309 Serial Numbers: 93202512, 93202514, 93202526, 93202527