Z-1394-2011
Maquet Cardiopulmonary AG, MAQUET RPM 20: Roller Pump for Heart Lung Machine HL20
Maquet Cardiovascular Us Sales, LLC / initiated 2010-09-28 / Terminated
Maquet Cardiovascular Us Sales, Llc began a recall of Maquet Cardiopulmonary AG, MAQUET RPM 20: Roller Pump for Heart Lung Machine HL20 on . The reason FDA records is "A wrong ball bearing was installed in the pump heads of product. This may make the pump head wear out and cease function". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1394-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Maquet Cardiopulmonary AG, MAQUET RPM 20: Roller Pump for Heart Lung Machine HL20.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- A wrong ball bearing was installed in the pump heads of product. This may make the pump head wear out and cease function.
- FDA root cause
- Component change control
- Action
- Maquet hand delivered a letter to the consignee on 10/4/2010.
- Quantity in commerce
- 4 units
- Distribution
- One hospital in New York City
- Product codes and lots
- Device Part Number (Model Number): 703309 Serial Numbers: 93202512, 93202514, 93202526, 93202527