Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump...
B Braun Medical, Inc began a recall of Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets on . The reason FDA records is "Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0837-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- B Braun Medical Inc
- Product
- Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets
- Product code
- FRN Pump, Infusion
- Reason for recall
- Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operation.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- B. Braun Medical Inc. notified consignee by letter on August 24, 2010.
- Quantity in commerce
- 2,107 Pumps (582 Pumps to Winchester)
- Distribution
- Virginia and New York
- Product codes and lots
- Model No. 621-400ES. Premarket Notifications K011975 and K994375. Device Listings D051037 and D051592.