Z-0837-2011

Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump...

B Braun Medical Inc / initiated 2010-08-24 / Terminated / root cause: design

B Braun Medical, Inc began a recall of Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets on . The reason FDA records is "Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0837-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
B Braun Medical Inc
Product
Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets
Product code
FRN Pump, Infusion
Reason for recall
Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operation.
Root cause tag
design
FDA root cause
Component design/selection
Action
B. Braun Medical Inc. notified consignee by letter on August 24, 2010.
Quantity in commerce
2,107 Pumps (582 Pumps to Winchester)
Distribution
Virginia and New York
Product codes and lots
Model No. 621-400ES. Premarket Notifications K011975 and K994375. Device Listings D051037 and D051592.