Z-0606-2011
Beaver-Visitec International (Beaver-Visitec) BD 1% OVD Sodium Hyaluronate, 10 mg/ml 1.0 mL. Sterile. Rx For...
Becton Dickinson & Co / initiated 2010-10-04 / Terminated / root cause: design
Becton Dickinson and Company began a recall of Beaver-Visitec International (Beaver-Visitec) BD 1% OVD Sodium Hyaluronate, 10 mg/ml 1.0 mL. Sterile. Rx For Intraocular Use REF 585304 Use as a surgical aid... on . The reason FDA records is "Finger flange on the BD 1% OVD syringe can disengage during cataract surgery and and may lead to injury". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0606-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton Dickinson & Co
- Product
- Beaver-Visitec International (Beaver-Visitec) BD 1% OVD Sodium Hyaluronate, 10 mg/ml 1.0 mL. Sterile. Rx For Intraocular Use REF 585304 Use as a surgical aid to protect corneal endothelium during cataract extraction procedures, intraocular lens implantation and anterior segment surgery.
- Product code
- LZP Aid, Surgical, Viscoelastic
- Reason for recall
- Finger flange on the BD 1% OVD syringe can disengage during cataract surgery and and may lead to injury.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Beaver Visitec issued an Urgent: Field Safety Notice letter, dated 10/4/10 to direct accounts, identifying the affected product and informing users of the potential for the finger grip to disengage during use. The firm offers solutions as follows: Allow the finger grip to remain attached to the syringe and continue to use the product. Position the open side of the finger grip towards the palm to avoid detachment of the grip from the syringe. Remove the finger grip and continue to use the product. Return the unused product to Beaver-Visitec for reimbursement. Users were requested to complete the Customer Response Form. Questions can be directed to: Inside Sales at 781-906-7952.
- Quantity in commerce
- 72278 units
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot No. Exp. Date RDOl71B 2012-12; RD0070A 2012-05; RD0018A 2012-02; RD0007A 2012-01 RC0064C 2011-06; RC0004A 2011-01;
FIRST ARTICLE / WEEKLY