Z-1519-2011

Pacific Hemostasis Thromboplastin-DS, catalog #100354, packaged in 4 mL lypholyzed vials, 10 bottles/carton...

Fisher Diagnostics / initiated 2010-10-06 / Terminated / root cause: manufacturing process

Fisher Diagnostics, A Company of Fisher Scientific LLC began a recall of Pacific Hemostasis Thromboplastin-DS, catalog #100354, packaged in 4 mL lypholyzed vials, 10 bottles/carton. Vials are labeled, in part ***Fisher Diagnostics a... on . The reason FDA records is "An in vitro diagnostic reagent used in coagulation testing may cause inaccurate clotting times in quality control and patient plasma samples". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1519-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fisher Diagnostics
Product
Pacific Hemostasis Thromboplastin-DS, catalog #100354, packaged in 4 mL lypholyzed vials, 10 bottles/carton. Vials are labeled, in part ***Fisher Diagnostics a division of Fisher Scientific Company, LLC Middletown, VA 22645-1905 USA*** Thromboplastin-DS is a Class 2 medical device intended for use in performing the one stage prothrombin time (PT) test and PT-based factor assays.
Product code
GJS Test, Time, Prothrombin
Reason for recall
An in vitro diagnostic reagent used in coagulation testing may cause inaccurate clotting times in quality control and patient plasma samples.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm, ThermoFisher Scientific, sent two "Urgent Medical Device Removal" letters one via Federal Express on October 20, 2010 and certified mail for domestic customers on October 21, 2010. The letters described the product, problem and action to be taken by the customers. The customers were instructed to indicate the quantities of all affected product currently in stock at their facility; complete and return the attached Response Form/table to Fisher Diagnostics via fax to (540) 869-8129 or by mail to the attention of Technical Support; dispose of any remaining inventory of the affected lot numbers; provide a copy of this letter to their customers, and file this letter for future reference. If you have any questions regarding this notification, please contact our Technical Support staff at 800-528-0494, Option 2, or email [email protected].
Quantity in commerce
1190 kits
Distribution
Worldwide distribution: USA including states of: Alaska, Florida, Kentucky, Ohio, New Jersey, New York, and Texas; and countries including: Argentina, Dubai, Italy, Mexico, Philippines, Romania, and Turkey.
Product codes and lots
V40526, exp 4/2013 V40826, exp 4/2013