Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0463-2011

3M ESPE Stainless Steel Primary Molar Crowns, Made in USA by 3M ESPE Dental Products, St Paul, MN. Unitek"...

3M Company / initiated 2010-10-06 / Terminated / root cause: packaging

3M Company / Medical Division began a recall of 3M ESPE Stainless Steel Primary Molar Crowns, Made in USA by 3M ESPE Dental Products, St Paul, MN. Unitek" Primary Stainless Steel Crowns- Series UL3 and UR3... on . The reason FDA records is "Unitek Stainless Steel Permanent Crowns with product Codes 902150, 900213, and 900223, is mislabeled. The crown contained within the package is incorrectly labeled; series UL3 crowns are labeled series UR3 and vice...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0463-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
3M Company
Product
3M ESPE Stainless Steel Primary Molar Crowns, Made in USA by 3M ESPE Dental Products, St Paul, MN. Unitek" Primary Stainless Steel Crowns- Series UL3 and UR3 Product Code 900223, 900213, 902150
Product code
ELZ Crown, Preformed
Reason for recall
Unitek Stainless Steel Permanent Crowns with product Codes 902150, 900213, and 900223, is mislabeled. The crown contained within the package is incorrectly labeled; series UL3 crowns are labeled series UR3 and vice versa. The lid of the package correctly identified the product number and lot. This mislabeling could result in placement on the right side of a crown intended for use on the left, and
Root cause tag
packaging
FDA root cause
Packaging process control
Action
3M ESPE issued a Device Recall letter dated October 6, 2010 to distributors and dentists. The letter listed the product being recalled including code and lot codes, background information including a picture of the product problem, and the recall procedure. The letter included a template letter to be used by distributors to inform their customers who have received the product. Mislabeled product is to be quarantined and returned to 3M ESPE. 3M ESPE can be contacted at 651 733-1110.
Quantity in commerce
9819
Distribution
Nationwide Distribution: Throughout USA, including the states of AZ, CA, GA, IL, IN, IA, LA, MA, MN, MO, NV,NJ, NY, NC, PA, TN, TX, WA, and WI. Canada and Dominican Republic
Product codes and lots
Product Code 900223 (Refill) " Lot codes N161092, N195115, N171401,N172028,N181008  Product Code 900213 (Refill) " Lot codes N161243, N168400, N181006, N186982, N195318  Product Code 902150 (Kit) " Lot Codes N155749, N170175, N193318