3M ESPE Stainless Steel Primary Molar Crowns, Made in USA by 3M ESPE Dental Products, St Paul, MN. Unitek"...
3M Company / Medical Division began a recall of 3M ESPE Stainless Steel Primary Molar Crowns, Made in USA by 3M ESPE Dental Products, St Paul, MN. Unitek" Primary Stainless Steel Crowns- Series UL3 and UR3... on . The reason FDA records is "Unitek Stainless Steel Permanent Crowns with product Codes 902150, 900213, and 900223, is mislabeled. The crown contained within the package is incorrectly labeled; series UL3 crowns are labeled series UR3 and vice...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0463-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- 3M Company
- Product
- 3M ESPE Stainless Steel Primary Molar Crowns, Made in USA by 3M ESPE Dental Products, St Paul, MN. Unitek" Primary Stainless Steel Crowns- Series UL3 and UR3 Product Code 900223, 900213, 902150
- Product code
- ELZ Crown, Preformed
- Reason for recall
- Unitek Stainless Steel Permanent Crowns with product Codes 902150, 900213, and 900223, is mislabeled. The crown contained within the package is incorrectly labeled; series UL3 crowns are labeled series UR3 and vice versa. The lid of the package correctly identified the product number and lot. This mislabeling could result in placement on the right side of a crown intended for use on the left, and
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- 3M ESPE issued a Device Recall letter dated October 6, 2010 to distributors and dentists. The letter listed the product being recalled including code and lot codes, background information including a picture of the product problem, and the recall procedure. The letter included a template letter to be used by distributors to inform their customers who have received the product. Mislabeled product is to be quarantined and returned to 3M ESPE. 3M ESPE can be contacted at 651 733-1110.
- Quantity in commerce
- 9819
- Distribution
- Nationwide Distribution: Throughout USA, including the states of AZ, CA, GA, IL, IN, IA, LA, MA, MN, MO, NV,NJ, NY, NC, PA, TN, TX, WA, and WI. Canada and Dominican Republic
- Product codes and lots
- Product Code 900223 (Refill) " Lot codes N161092, N195115, N171401,N172028,N181008 Product Code 900213 (Refill) " Lot codes N161243, N168400, N181006, N186982, N195318 Product Code 902150 (Kit) " Lot Codes N155749, N170175, N193318