Z-0542-2011
Edwards Protection Cannulae - Monitoring and Infusion Set, Model ISP3147, Sterile EO, Edwards Lifesciences...
Edwards Lifesciences LLC / initiated 2010-10-19 / Terminated / root cause: manufacturing process
Edwards Lifesciences Llc began a recall of Edwards Protection Cannulae - Monitoring and Infusion Set, Model ISP3147, Sterile EO, Edwards Lifesciences LLC, One Edwards Way, Irvine, CA 92614 on . The reason FDA records is "Monitoring and Infusion Set connectors may break during manipulation". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0542-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Edwards Lifesciences LLC
- Product
- Edwards Protection Cannulae - Monitoring and Infusion Set, Model ISP3147, Sterile EO, Edwards Lifesciences LLC, One Edwards Way, Irvine, CA 92614
- Product code
- DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
- Reason for recall
- Monitoring and Infusion Set connectors may break during manipulation.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Edwards Lifesciences issued a Urgent Product Recall letter dated October 13, 2010 to customers identifying the affected devices, requesting the return of unused devices and completion and return of a response form. Further information is available at 800-424-3278.
- Quantity in commerce
- 322 sets
- Distribution
- Nationwide Distribution: in the states of FL, KS, MO, TX, VA, and PR.
- Product codes and lots
- Lot No.: 58803802, 58851054, 58881366