Z-0428-2011

Penumbra System Reperfusion Catheter 032, catalog number PSC032, lot number lot F15020, manufactured by...

Penumbra Inc. / initiated 2010-10-25 / Terminated / root cause: manufacturing process

Penumbra Inc. began a recall of Penumbra System Reperfusion Catheter 032, catalog number PSC032, lot number lot F15020, manufactured by Penumbra, Alameda, CA on . The reason FDA records is "Reperfusion catheter may fail at the mid-shaft joint and may break off. If the segment cannot be removed, it may result in blocked blood flow, requiring emergency medical treatment". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0428-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Penumbra Inc.
Product
Penumbra System Reperfusion Catheter 032, catalog number PSC032, lot number lot F15020, manufactured by Penumbra, Alameda, CA.
Product code
NRY Catheter, Thrombus Retriever
Reason for recall
Reperfusion catheter may fail at the mid-shaft joint and may break off. If the segment cannot be removed, it may result in blocked blood flow, requiring emergency medical treatment.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Penumbra issued an Urgent Voluntary Field Removal letter dated October 28, 2010 to consignees, identifying the affected device and actions to be taken by consignees. Customers were instructed to inspect their inventory for the affected product and quarantine it, pending return to Penumbra. Penumbra employees will contact customers to arrange return and replacement of affected units. Customers were also requested to complete and return a product identification form and return form to Penumbra within 5 business days. Customers can contact Penumbra at 1 510 748-3223 concerning this action.
Quantity in commerce
61
Distribution
Worldwide Distribution: 17 consignees in the US, including in the states of CA, FL, HI, ID, IL, KY, MA, MS, NJ, NY, PA, and TX, 1 in Canada and 21 in the EU countries of France, Germany, Hungary, Italy, Norway, Slovenia, Spain, and the United Kingdom.
Product codes and lots
Lot number F15020