Z-0557-2011

Custom Kit, Sterile EO, Rx Only, Catalog No. K09-09815, Manufacturer: Merit Medical Systems, Inc. South...

Merit Medical Systems, Inc. / initiated 2010-10-25 / Terminated / root cause: manufacturing process

Merit Medical Systems, Inc. began a recall of Custom Kit, Sterile EO, Rx Only, Catalog No. K09-09815, Manufacturer: Merit Medical Systems, Inc. South Jordan, Utah 84095, USA on . The reason FDA records is "Defective custom kits may draw in air during use which may compromise patient safety and affect treatment. The valve's fixed luer may be over- or under-tightened, causing leaking at the bond joint". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0557-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
Custom Kit, Sterile EO, Rx Only, Catalog No. K09-09815, Manufacturer: Merit Medical Systems, Inc. South Jordan, Utah 84095, USA,
Product code
DQO Catheter, Intravascular, Diagnostic
Reason for recall
Defective custom kits may draw in air during use which may compromise patient safety and affect treatment. The valve's fixed luer may be over- or under-tightened, causing leaking at the bond joint.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Merit issued an Urgent Product Recall Notice letter dated October 25, 2010 to consignees. Copnsignees were instructed to quarantine the product immediately for return and to notify customers of recall activity and instruct them to return product.. All distributors and customers are to complete and return a Recall Notification Form. For further information about this recall, contact Merit at 800 356-3748.
Quantity in commerce
1033 units
Distribution
Nationwide Distribution in the states of CA, IL, IN, and UT.
Product codes and lots
Lot No.: H137939, H147372, H151609, H158544