Z-0532-2011

Zimmer, NexGen Complete Knee Solution LPS Femoral Component, Size E Right, sterile, REF 00-5996-015-02

Zimmer GmbH / initiated 2010-09-15 / Terminated / root cause: manufacturing process

Zimmer Inc. began a recall of Zimmer, NexGen Complete Knee Solution LPS Femoral Component, Size E Right, sterile, REF 00-5996-015-02 on . The reason FDA records is "These lots of Zimmer NexGen Complete Knee Solution LPS Femoral Component and Zimmer NexGen Complete Knee Solution LPS Flex Gender Femoral Component exhibit a nonconforming internal CAM radius. The condition is...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0532-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Zimmer, NexGen Complete Knee Solution LPS Femoral Component, Size E Right, sterile, REF 00-5996-015-02.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
These lots of Zimmer NexGen Complete Knee Solution LPS Femoral Component and Zimmer NexGen Complete Knee Solution LPS Flex Gender Femoral Component exhibit a nonconforming internal CAM radius. The condition is intermittent and not all devices have nonconforming geometry.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The firm sent URGENT DEVICE CORRECTION AND REMOVAL notices to distributors and sales staff, Risk Managers, and implanting surgeons dated 9/15/2010. Sales staff and risk managers were instructed to return any product on hand. Surgeons were informed of the issue but there was no action to be taken by surgeons.
Quantity in commerce
192 all sizes of 5996.
Distribution
US, Germany, Spain, France, UK, Italy, Sweden, South Africa, Czech Republic, Bulgaria, Poland, Romania, and Austria.
Product codes and lots
61494589