Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0945-2011

GE Healthcare Discovery RX (16 slice) PET/CT System; Discovery RX (16 Slice) PET/CT Systems with technical...

GE Healthcare / initiated 2010-10-27 / Terminated

GE Healthcare, LLC began a recall of GE Healthcare Discovery RX (16 slice) PET/CT System; Discovery RX (16 Slice) PET/CT Systems with technical Reference Manual numbers 2351785-xxx and 2390935-xxx... on . The reason FDA records is "Although the system x-ray output is accurate for the technique factors given, and the dose display is also accurate, certain CTDI adjustment factors used for the manual calculation from the Technical Reference Manual...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0945-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare Discovery RX (16 slice) PET/CT System; Discovery RX (16 Slice) PET/CT Systems with technical Reference Manual numbers 2351785-xxx and 2390935-xxx distributed from November 2001 to August 2010. Head and whole body X-ray computed tomography applications.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
Although the system x-ray output is accurate for the technique factors given, and the dose display is also accurate, certain CTDI adjustment factors used for the manual calculation from the Technical Reference Manual (TRM) are incorrect. CTDI100 Aperture Adjustment Factors for 4x0.625 mode contain incorrect values. This issue may lead to inaccurate CTDI manual calculation. The results of the diffe
FDA root cause
Radiation Control for Health and Safety Act
Action
An Important Electronic Product Radiation Warning letter will be provided together with the TRM addendum informing the customers of the issue and related hazards. The letter also identified the affected product and gave instructions, which include referring to the TRM addendum, along with the product correction. This correction will be executed via a Field Modification Instruction IFMI 254281 released in November, 2010 with an estimated completion date of May, 2011. Questions or concerns should be directed to the call numbers listed or a local GE Healthcare Service Representative for countries not listed.
Quantity in commerce
1450 units total in US
Distribution
Worldwide Distribution
Product codes and lots
Technical Reference Manual numbers 2351785-xxx and 2390935-xxx.