Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1674-2011

BLACKSTONE Medical Inc. Firebird Spinal Fixation System 6.5mm x 30mm Bone Screw, Self Tapping, Cannulated...

Orthofix SRL / initiated 2010-10-22 / Terminated

Orthofix Inc began a recall of BLACKSTONE Medical Inc. Firebird Spinal Fixation System 6.5mm x 30mm Bone Screw, Self Tapping, Cannulated. BLACKSTONE Medical Inc. Firebird Spinal Fixation... on . The reason FDA records is "There is a possibility that certain Multi-Axial Screws have spherical heads measuring less than the allowable tolerance". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1674-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
BLACKSTONE Medical Inc. Firebird Spinal Fixation System 6.5mm x 30mm Bone Screw, Self Tapping, Cannulated. BLACKSTONE Medical Inc. Firebird Spinal Fixation System 6.5mm x 30mm Multi-Axial Screw, Self Tapping, Cannulated. The Multi-Axial Screw is a pedicle screw which is inserted into the pedicle with the use of the Multi-Axial Screw Driver. The component is used within the Firebird Spinal Fixation System. The screw is cannulated and available modular (77-7630, pre-assembled with screw head) or non-modular (77-8630, without screw head). Product Code: NKB-Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease. Class II. Intended for use when implanted over a guide wire for additional accuracy with reference to the pedicle.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
There is a possibility that certain Multi-Axial Screws have spherical heads measuring less than the allowable tolerance.
FDA root cause
Other
Action
Orthofix sent an URGENT: PRODUCT RECALL letter dated October 22, 2010. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to contact their accounts immediately and advise them of the recall situation and have them return any affected inventory. For any questions customers were instructed to contact their Account Services Representative at 1-888-298-5700. For questions regarding this recall call 214-937-2061.
Quantity in commerce
27 (19 of 77-7630) (8 of 77-8630)
Distribution
Nationwide Distribution including AZ, NY, TX, MS, LA, FL, NC, and AR.
Product codes and lots
Catalog #'s: 77-7630 and 77-8630.  510(k) Information: K081684 (Blackstone Medical Inc.). Lot #: R01. UPC's: M3327776301G and M3327786301H.