Z-0918-2011

Philips OmniDiagnost Eleva with Digital Image (DI) Rel.2 system. The Philips OmniDiagnost Eleva is a...

Philips North America LLC / initiated 2009-03-06 / Terminated

Philips Medical Systems North America Co. Phillips began a recall of Philips OmniDiagnost Eleva with Digital Image (DI) Rel.2 system. The Philips OmniDiagnost Eleva is a diagnostic-imaging device intended for radiographic... on . The reason FDA records is "Images from two different patients may be combined in one patient folder in the Digital Image (DI) or may be combined as one (printed) image. It is possible that an image of a previous patient registered in the DI will...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0918-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips OmniDiagnost Eleva with Digital Image (DI) Rel.2 system. The Philips OmniDiagnost Eleva is a diagnostic-imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
Images from two different patients may be combined in one patient folder in the Digital Image (DI) or may be combined as one (printed) image. It is possible that an image of a previous patient registered in the DI will be overwritten with an image of a later patient.
FDA root cause
Other
Action
On March 6, 2009, Philips sent an "URGENT-Device Correction Easy Diagnost Eleva, MultiDiagnost Eleva, and OmniDiagnost Eleva with DI Rel. 2 systems notification" to their consignees/customers. The consignees/customers are advised that before diagnosing images, always check the stored (Digital Image, Viewform or PACS) or printed images to ensure that these represent the data and images captured from the examination of the respective patient. Do not use images if they do not represent the patient that was examined. Consignees will be contacted by Philips to schedule a software upgrade of their units. The consignees can contact Philips Healthcare Call Center at 800-722-9377, #5, #2 and reference "FCO 70800070" with any questions about this recall.
Quantity in commerce
13 units
Distribution
Nationwide distribution: USA including states of: AZ, CA, CO, CT, DE, FL, GA, HI, ID, LA, MA, ME, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OR, PA, SC, TN, TX, VA, WA, WI, and WY.
Product codes and lots
Site Numbers:   533530, 535118, 535257, 535924, 539106, 540860, 541490, 547017, 547142, 547348, 549798, 550258, and 553235.