Z-0591-2011
CHROMOPHARE Surgical Lighting System & CHROMOVIEW Flat Panel Monitor, Mechanical Spring Arms
Berchtold Corp. / initiated 2009-11-16 / Terminated / root cause: manufacturing process
Berchtold Corp. began a recall of CHROMOPHARE Surgical Lighting System & CHROMOVIEW Flat Panel Monitor, Mechanical Spring Arms on . The reason FDA records is "A mechanical failure of a welded joint within the spring Arm may result in the rapid drop of a monitor to the mechanical stop". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0591-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Berchtold Corp.
- Product
- CHROMOPHARE Surgical Lighting System & CHROMOVIEW Flat Panel Monitor, Mechanical Spring Arms.
- Product code
- FTD Lamp, Surgical
- Reason for recall
- A mechanical failure of a welded joint within the spring Arm may result in the rapid drop of a monitor to the mechanical stop.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Berchtold issued a Product Safety Notification letter on November 16, 2009 identifying the affected device and issue prompting the action. Consignees were instructed to: Exercise caution when adjusting the position of the surgical lights or flat panel monitors. While an event is highly unlikely, the likelihood does increase when the spring arm is being repositioned. Identify a specific contact person for their facility to facilitate the communication and field action process with BERCHTOLD. The following action is being planned: To further identify if their units are part of the affected population, determine the supplier serial number identification and specific location for each spring arm in their facility. Report the facility information, supplier spring arm serial numbers and specific location or spring arms to a BERCHTOLD. Detail instructions on serial number identification and the method of reporting to BERCHTOLD will also be provided in a subsequent communication. Once BERCHTOLD has received the serial numbers of their spring arm assembly, customers will be notified whether their particular units are affected and the further required action. Berchtold Corporation can be contacted at 1 800 968 6097 about this action.
- Quantity in commerce
- 7031 arms
- Distribution
- Worldwide Distribution: Throught the U.S,.and the countries of Canada and Germany
- Product codes and lots
- Spring Arm Serial Numbers: 010103837915-38114, 010101835874-35973, 010101836138, 010103837680-37779, 010101836309-36408, 10109841913-41942, 010107841041-41060, 010106840378-40407, 010104838741-38770, 010102837580-37609, 010102837341-37370, 01010183669, 010112735745-35804, 010112735193-35302, 010111734501-34663, 010103838406-38555, 010101835988-36137, 010103838574-38603, 010102837465-37494, 010102837301-37340, 010112735125-35134, 010109946662-46711, 010106945840-45866, 010110842349-42388, 010107841061-41100, 010107840798-40837, 010107840778-40797, 010103838286-38325, 010103838286-38325, 010102836829-36868, 010109946642-46661, 010107946042-46050, 010109842141-42180, 010109842140-42140, 010109841871-41910, 010108841499-41538, 010107840959-40998, 010106840715-40754, 010104839108-39147, 010105839565-39604, 010104839098-39107, 010104838907-38961, 010102837525-37578, 010102837579, 010101836778-36813, 010101835974-35977, 010204803835-03844, 010202802298-02307, 010202802223-02232, 010201801889-01898, 010201801569-01578, 010211807820-07969, 010211807720-07819, 010210807565-07614, 010210807454-07553, 010210807254-07453, 010209806644-06893, 010209806504-06603, 010208806352-06501, 010208806262-06311, 010208806092-06141, 010207805722-05771, 010207805479-05681, 010206805229-05328, 010206805079-05228, 010205804650-04749, 010205804358-04407, 010204803965-04164, 010204803965-04014, 010204803535-03634, 010202802579-02628, 010201802039-02088, 010201801839-01888, 010201801189-01338, 010205804500-04649, 010204804220-04269, 010205804408-04457, 010204804170-04219, 010204804065-04164, 010204803635-03734, 010203803485-03534, 010203803375-03424, 010203803275-03374, 010203802709-02808, 010202802308-02407, 010202802408-02507, 010201801899-01998, 010201801639-01738, 010201801739-01838, 010201801089-01188, 010212808270-08409, 010211808210-08269, 010209806994-07053, 010209806894-06993, 010209806604-06643, 010208806312-06351, 010208806202-06261, 010208806142-06201, 010207805979-06038, 010207805772-05871, 010204803865-03964, 010104803735-03834, 010203802909-03008, 010201802089-02188, 010201801999-02038, 010201801579-01638, 010207805892-05911, 010207805872-05891 and 010105839563-39564.