Z-0797-2011

Vital, Vitrea fX and Vitrea Enterprise Suite fX Model number: Vitrea fX 2.1, fX 3.0, and fX 3.1 Vitrea...

Canon Medical Informatics, Inc. / initiated 2010-10-14 / Terminated / root cause: software

Vital Images, Inc. began a recall of Vital, Vitrea fX and Vitrea Enterprise Suite fX Model number: Vitrea fX 2.1, fX 3.0, and fX 3.1 Vitrea Enterprise Suite (VES) fX 1.2 and VES fX 1.3 The Vitrea... on . The reason FDA records is "Vital Images Inc., reported that a software defect exists with Vitrea fX 2.1, 3.0 and 3.1 CT Brain Analysis whereby the annotated numeric ROI values on a batched perfusion map are incorrect when created using the auto...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0797-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Canon Medical Informatics, Inc.
Product
Vital, Vitrea fX and Vitrea Enterprise Suite fX Model number: Vitrea fX 2.1, fX 3.0, and fX 3.1 Vitrea Enterprise Suite (VES) fX 1.2 and VES fX 1.3 The Vitrea 4D CT Brain Perfusion option is intended for post processing based on dynamic CT images continuously acquired during the infection of contrast, for the visualization of apparent blood flow in brain tissue and pictorial illustration of perfusion-related parameters to aid in the assessment of the type and extent of cerebral perfusion disturbances.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Vital Images Inc., reported that a software defect exists with Vitrea fX 2.1, 3.0 and 3.1 CT Brain Analysis whereby the annotated numeric ROI values on a batched perfusion map are incorrect when created using the auto batching function (from the right click menu) with the auto ROI templates enabled.
Root cause tag
software
FDA root cause
Software design
Action
Vital Images, Inc sent a Vital "Urgent Software Update Alert" letter dated 14 October 2010 to all Consignees/Customers. The letter described the product and problem. They advised, until new software is installed, that the customers can avoid obtaining incorrect values by tuning off ROI templates prior to selecting the batch option from the right click menu in the CT Brain Analysis application; and recommended that the customers re-examine any ROI perfusion values that were previously generated using automatic collage batching with ROI templates. Please direct any questions or comments to Customer Support at [email protected] or at 1-800-208-3005. Any help needed to install this software update can also be directed to Customer Support.
Quantity in commerce
182
Distribution
Worldwide distribution: USA including states of: AL, AZ, AR, CA, CO, FL, GA, IA, KY, LA, MD, MA, MI, MN, MO, MT, NY, NJ, ND, OH, PA, PR, TN, TX, WA, WI, and WY, and countries of: CANADA, NETHERLANDS, TAIWAN, CHINA, SINGAPORE, UNITED KINGDOM, BRAZIL, GERMANY, RUSSIA, THAILAND, FRANCE, REPUBLIC OF KOREA, SPAIN, PAKISTAN,IRELAND, JAPAN, AUSTRALIA, INDIA, CZECH REPUBLIC, HONG KONG, EGYPT, ITALY, AUSTRIA, SLOVAK IA, BULGARIA, LEBANON, ALGERA, TAIWAN, SOUTH ARICA, BELGIUM, MALAYSIA, DENMARK, and JORDAN.CANADA, NETHERLANDS, TAIWAN, CHINA, SINGAPORE, UNITED KINGDOM, BRAZIL, GERMANY, RUSSIA, THAILAND, FRANCE, REPUBLIC OF KOREA, SPAIN, PAKISTAN,IRELAND, JAPAN, AUSTRALIA, INDIA, CZECH REPUBLIC, HONG KONG, EGYPT, ITALY, AUSTRIA, SLOVAK IA, BULGARIA, LEBANON, ALGERA, TAIWAN, SOUTH ARICA, BELGIUM, MALAYSIA, DENMARK, JORDAN.
Product codes and lots