Z-1389-2011

S-ROM Distal Femoral Cutting block Guide SZ MED, REF 216311004, non-sterile, Depuy International Leeds...

Depuy Orthopaedics Inc. / initiated 2010-10-19 / Terminated / root cause: design

Depuy Orthopaedics, Inc. began a recall of S-ROM Distal Femoral Cutting block Guide SZ MED, REF 216311004, non-sterile, Depuy International Leeds, England. The product is used to make cuts in the femur... on . The reason FDA records is "The firm was notified that the Hinge Block Femoral Box Guide cut being off alignment when using the S-ROM Distal Femoral Cutting Block Guide SZ SML/X-SML pin holes to locate. The firm then found that the same issue was...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1389-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Depuy Orthopaedics Inc.
Product
S-ROM Distal Femoral Cutting block Guide SZ MED, REF 216311004, non-sterile, Depuy International Leeds, England. The product is used to make cuts in the femur to prepare the femur for total knee replacement surgery.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The firm was notified that the Hinge Block Femoral Box Guide cut being off alignment when using the S-ROM Distal Femoral Cutting Block Guide SZ SML/X-SML pin holes to locate. The firm then found that the same issue was found int the MED block size.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, DePuy, sent an "URGENT INFORMATION - DEVICE CORRECTION" email and letter dated October 19, 2010 to all customers. The notices described the product, problem and actions to be taken by the customers. The notices stated that: Depuy is issuing a device correction; however, the product may still be used; the devices do not need to be returned; the distributors were to provide a letter to clinicians that were using the device; the surgical technique will be revised in October 2010; the product will undergo a design change, and the recalled product will be traded out in December 2010 and January 2011. If you have any additional questions, please contact DePuy's Scientific Information office at 1-888-554-2482. On 3/4/2011, The firm sent out a revised URGENT DEVICE RECALL asking that all affected cutting guides be taken out of service immediately. Devices should be returned to DePuy and new cutting guides will be sent out within 2 weeks.
Quantity in commerce
219
Distribution
Worldwide distribution: USA including states of: AZ, AR, CA, CO, FL, GA, IA, ID,IL, IN, KY, KS, LA, MA, MD, MI, MN, NC, NV, PA, TN, TX, VA, WA, and WI; and countries including: Canada and Ireland. Australia, Austria, Canada, Denmark, Ireland, Israel Italy, New Zealand, Russia, Sweden, Switzerland, Spain, and UK.
Product codes and lots
C44AK4000, D1WL94000, DG9LA4000, DJ5EF4000, DJ5EV4000, DY6GB4000 and DP5WM4000.