GE Healthcare Discovery MR750 3.0T The GE Signa MR750 System is a whole body magnetic resonance scanner...
GE Healthcare, LLC began a recall of GE Healthcare Discovery MR750 3.0T The GE Signa MR750 System is a whole body magnetic resonance scanner designed to support high resolution, high... on . The reason FDA records is "GE Healthcare has become aware that during certain service procedures, there is a potential for electric shock associated with the gradient cooling manifold hoses of your MR750, MR450, and MR450w MR systems that may...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1375-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Discovery MR750 3.0T The GE Signa MR750 System is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The GE Signa MR750 System is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and organs of the entire body, including, but not limited to, head, neck, TMVJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the GE Signa MR750 System reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images when interpreted by a trained physician yield information that may assist in diagnosis. The indications for use for the 3.0T GE Signa MR750 System are similar to those for the Signa HDx Magnetic Resonance System and the Siemens MAGNETOM Verio MR System.
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- GE Healthcare has become aware that during certain service procedures, there is a potential for electric shock associated with the gradient cooling manifold hoses of your MR750, MR450, and MR450w MR systems that may impact service or maintenance personnels safety. The gradient cooling manifolds used on the affected systems may contain electrically conductive hoses, which become live when the g
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- GE Healthcare sent an "Urgent Medical Device Correction" letter dated November 16, 2011, to all affected customers. The letter was addressed to Hospital Administrator / Risk Manger, Radiology Department Manager and Maintenance / Biomedical Engineering Department. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product correction and Contact Information. Customers were instructed to follow appropriate lockout/tagout (LOTO) procedures. Customers were instructed to contact GE Healthcare engineering at 843-687-7213 or 843-697-3009 for instructions on how to proceed. For questions concerning the notiifcaton customers were instructed to contact their local Support Engineer.
- Quantity in commerce
- 25
- Distribution
- Worldwide Distribution - USA including AL, AZ, CA, CO, CT, DE, DC, FL, GA, IL, IN, IA, LA, MD, MA, MI, MO, NV, NJ, NY, NC, OR, PA, SC, TN, TX, WA, WI. and the countries of AUSTRALIA, BRAZIL, CANADA, EHYPT, FINLAND, FRANCE, GERMANY, GREECE, INDIA, ITALY, JAPAN, KUWAIT, QATAR, REPUBLIC OF KOREA, SWITZERLAND, TAIWAN, and the UNITED KINGDOM.UNITED KINGDOM, TAIWAN, SWITZERLAND, REPUBLIC OF KOREA, QATAR, KUWAIT, JAPAN, ITALY, INDIA, GREECE, GERMANY, FRANCE, FINLAND, EGYPT, CANADA, BRAZIL, AUSTRALIA.
- Product codes and lots
- 00000XRMBB0092 00000XRMBB0093 00000XRMBB0098 00000XRMBB0102 00000XRMBB0103 00000XRMBB0106 00000XRMBB0108 00000XRMBB0113 00000XRMBB0115 00000XRMBB0117 00000XRMBB0128 00000XRMBB0134 00000XRMBB0112 00000XRMBB0107 00000XRMBB0076 00000XRMBB0061 00000XRMBB0116 00000XRMBB0077 00000XRMBB0097 00000XRMBB0062 00000XRMBB0130 00000XRMBB0118 00000XRMBB0124 00000XRMBB0095 00000XRMBB0105