AngioScore, Inc. AngioSculpt PTA Scoring Balloon Catheter; OTW 0.018" Platform, 137 cm, 4.0 mm x 20 mm...
AngioScore Inc. began a recall of AngioScore, Inc. AngioSculpt PTA Scoring Balloon Catheter; OTW 0.018" Platform, 137 cm, 4.0 mm x 20 mm balloon size. Part (REF) number: 2076-4020 on . The reason FDA records is "Reports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end of the scoring element. May result in vessel injury or embolism". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0838-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Spectranetics Corp.
- Product
- AngioScore, Inc. AngioSculpt PTA Scoring Balloon Catheter; OTW 0.018" Platform, 137 cm, 4.0 mm x 20 mm balloon size. Part (REF) number: 2076-4020
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- Reports of distal bond failure which has resulted in peeling of the bond and/or detachment of the distal end of the scoring element. May result in vessel injury or embolism.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Medical Device Recall notifications were sent to all domestic consignees on Nov 16, 2010 by overnight mail. Foreign distributors, other than Japan, were sent recall letters by e-mail or international overnight delivery on November 16, 2010. The Japanese distributor received notification and notified their customers on November 10, 2010.
- Quantity in commerce
- 17682 total units, all sizes and lots
- Distribution
- Worldwide Distribution: USA, China, Finland, Italy, Japan, Norway, Saudi Arabia, Spain, Sweden, Switzerland, UAE, United Kingdom
- Product codes and lots
- Part (REF) number: 2076-4020: All lot codes