Z-0804-2011
ACL 9000 Coagulation Analyzer, PN 00000990011 Coagulation testing
Instrumentation Laboratory Co. / initiated 2010-11-11 / Terminated / root cause: design
Instrumentation Laboratory Co. began a recall of ACL 9000 Coagulation Analyzer, PN 00000990011 Coagulation testing on . The reason FDA records is "Sporadically lowered APTT results were observed with the PT/APTT Test Groups". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0804-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Instrumentation Laboratory Co.
- Product
- ACL 9000 Coagulation Analyzer, PN 00000990011 Coagulation testing.
- Product code
- JPA System, Multipurpose For In Vitro Coagulation Studies
- Reason for recall
- Sporadically lowered APTT results were observed with the PT/APTT Test Groups.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 11/11/10, An Urgent Product Notification informing customers to discontinue use of all IL locked PT/APTT-based Test Groups was mailed to U.S. and Canadian customers, instructing them to run in Single Test or Profile Mode until further notice. The letter identified the affected product, the issue, risk and stated a list of mandatory actions. Customers are to immediately acknowledge the receipt of the notification by signing and returning it via email or fax. A Worldwide Regulatory Notice was sent on November 8, 2010 to international affiliates (non-U.S. or Canada), notifying them of the issue, providing a risk summary and the support documentation for the notification. Secondary international dealers were notified Nov 8 and a fax response of notification of the dealers will be tracked by International Laboratories. The customer notification of the secondary dealers will be handled per the local regulatory requirements of the countries where they are located.
- Quantity in commerce
- 87 distributed in USA & Canada
- Distribution
- Worldwide Distribution -- USA, Canada, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Thailand, Trinidad & Tobago, Tunisia, Turkey, Turkmenistan, United Arabian Emirates, Uruguay, Venezuela, Vietnam, Austria, Belgium, China, France, Germany, Holland, Italy, Japan, Spain (and Portugal), and UK.
- Product codes and lots
- Serial numbers range 00040105 - 09031882.