Z-0806-2011

ACL ELITE Coagulation Analyzer, PN 00000880311 Coagulation testing

Instrumentation Laboratory Co. / initiated 2010-11-11 / Terminated / root cause: design

Instrumentation Laboratory Co. began a recall of ACL ELITE Coagulation Analyzer, PN 00000880311 Coagulation testing on . The reason FDA records is "Sporadically lowered APTT results were observed with the PT/APTT Test Groups". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0806-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Instrumentation Laboratory Co.
Product
ACL ELITE Coagulation Analyzer, PN 00000880311 Coagulation testing.
Product code
JPA System, Multipurpose For In Vitro Coagulation Studies
Reason for recall
Sporadically lowered APTT results were observed with the PT/APTT Test Groups.
Root cause tag
design
FDA root cause
Device Design
Action
On 11/11/10, An Urgent Product Notification informing customers to discontinue use of all IL locked PT/APTT-based Test Groups was mailed to U.S. and Canadian customers, instructing them to run in Single Test or Profile Mode until further notice. The letter identified the affected product, the issue, risk and stated a list of mandatory actions. Customers are to immediately acknowledge the receipt of the notification by signing and returning it via email or fax. A Worldwide Regulatory Notice was sent on November 8, 2010 to international affiliates (non-U.S. or Canada), notifying them of the issue, providing a risk summary and the support documentation for the notification. Secondary international dealers were notified Nov 8 and a fax response of notification of the dealers will be tracked by International Laboratories. The customer notification of the secondary dealers will be handled per the local regulatory requirements of the countries where they are located.
Quantity in commerce
514 distributed in USA & Canada, 55 in distribution channel subject to recall
Distribution
Worldwide Distribution -- USA, Canada, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Thailand, Trinidad & Tobago, Tunisia, Turkey, Turkmenistan, United Arabian Emirates, Uruguay, Venezuela, Vietnam, Austria, Belgium, China, France, Germany, Holland, Italy, Japan, Spain (and Portugal), and UK.
Product codes and lots
Serial numbers range 05010100 - 10111176.