Z-0715-2011
Polaris Spectra System Control Unit, a component of the StealthStation S7 System. Medtronic Navigation...
Medtronic Navigation, Inc / initiated 2010-11-05 / Terminated / root cause: design
Medtronic Navigation, Inc began a recall of Polaris Spectra System Control Unit, a component of the StealthStation S7 System. Medtronic Navigation, Louisville, CO 80027. Intended as an aid for precisely... on . The reason FDA records is "Electronic surgical navigation aid may lose tracking ability or stop working". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0715-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Navigation, Inc
- Product
- Polaris Spectra System Control Unit, a component of the StealthStation S7 System. Medtronic Navigation, Louisville, CO 80027. Intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures.
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Electronic surgical navigation aid may lose tracking ability or stop working.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Customers were notified by a Product Correction Notification letter on 11/05/2010. The letter advised customers of the affected product, the potential problem, and the description of the correction. Customers were told that a Medtronic rep will visit them to install a replacement control unit. Further information is available at 800-595-9709 or 720-890-3200.
- Quantity in commerce
- 48 devices
- Distribution
- Worldwide Distribution -- US, including states of AL, AZ, CO, FL, GA, KY, MD, MN, NE, PA, TX, WI, and MD and countries of Australia, Belgium, Canada, Czech Republic, Italy, Netherlands, Poland, Portugal, Saudi Arabia, Singapore, South Korea, Spain, Turkey, and UAE.
- Product codes and lots
- Lots received from NDI (supplier) between May 1, 2020, and July 31, 2010.